Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 125
Example: 0 ... 5 V voltage output selected for channel 1 and 4...20 mA
selected for channel 2.
OFF ON
Selection
1
█
2
█
Voltage output selected
3
█
4
█
5
█
6
█
0 ... 5 V selected
7
█
8
█
1
█
Current output selected
2
█
3
█
4
█
4 ... 20 mA selected
5
█
6
█
7
█
8
█
NOTE
If you have customized the error output setting (
AERR
), check that the
set error values are still valid after changing the output mode/range. See
section Analog Output Fault Indication Setting on page 128.
Analog Output Quantities
Use the display/keypad to change and scale the analog output quantities.
1.
Press any of the arrow buttons to open the
Main Menu.
2. Select
Interfaces
by pressing the
►
arrow button.
3. Select
Analog outputs
by pressing the
►
arrow button.
4. Select
Output 1/2/3
by pressing the
►
arrow button.
5. Select
Quantity
by pressing the
▲▼
arrow buttons. Confirm your
selection by pressing
CHANGE
.
6. Select
the quantity
by using the arrow buttons. Press
SELECT
to
confirm your selection.
7. Select
Scale,
lower limit,
by pressing the
▲▼
arrow buttons. Press
SET
to confirm your selection. Press
OK
to confirm your setting.