Analog Output Fault Indication Setting
Factory default state for analog outputs during error condition is 0 V/
0mA. Please be careful when selecting the new error value. The error
state of the transmitter should not cause unexpected problems in process
monitoring.
Use the display/keypad to set the analog output fault indication.
1.
Press any of the arrow buttons to open the
Main Menu.
2. Select
Interfaces
by pressing the
►
arrow button.
3. Select
Analog Outputs
by pressing the
►
arrow button.
4. Select
Output 1/2/3
by pressing the
►
arrow button.
5.
Select Fault indication. Press
SET
to confirm your selection. Enter
the fault indication value by using the arrow buttons. Press
OK
to
confirm your setting. This value is outputted if a transmitter error
occurs.
6. Press
EXIT
to return to the basic display.
AERR
Use the serial line
AERR
command to change the error output.
AERR
<cr>
Example:
>
aerr
Ch1 error out : 0.000V ?
5.0
Ch2 error out : 0.000V ?
5.0
>
NOTE
The error output value must be within a valid range for the output type.
NOTE
The error output value is displayed only when there are minor electrical
faults such as a humidity sensor damage. When there is a severe device
malfunction, the error output value is not necessarily outputted.