4.0 Setup Instructions and Installation
4.1
Components
When unpacking the GelDoc-It
TS3
and ChemiDoc-It
TS3
, the following items will be included:
GelDoc-It
TS3
or ChemiDoc-It
TS3
system
Camera, lens, mount, brackets and hardware
Stylus
Ethidium bromide (EtBr) emission filter
Transilluminator
Power cables
USB flash drive (8 GB minimum)
Keyboard and mouse
VisionWorks Software on a flash drive (if included; varies by country)
Supporting documentation
Black chemi tray for blots (ChemiDoc-It TS3 only)
Gel tray and gel ruler (GelDoc-It TS3 only)
When unpacking and setting up the darkroom, two people are required to move the darkroom.
Place the darkroom on a flat surface which can provide adequate support for up to 100 pounds.
WARNING:
Do not attempt to perform any setup procedures while the system is
plugged in or powered on unless otherwise instructed.
Do not install the system in areas with high moisture, dust or high temperatures. Keep
the equipment away from motors or any other large magnetic equipment. This system is
designed for indoor use only.
Do not position the system so that it is difficult to access the power cable and operate
the main power switch at the back of the unit.
4.2
Connecting the Transilluminator and Accessories
1.
Inside the darkroom, place the transilluminator on the roll-out tray. Connect the
transilluminator to the internal power jumper cable. Ensure that the green power switch on
the front of the transilluminator unit is in the
ON
position and that the desired wavelength or
intensity is selected.
Note
: For UV protection and to extend the life of the UV transilluminator, the system
incorporates a customizable transilluminator shutoff timer built into the software. For
additional information, refer to the
Lighting
section of this manual.
2.
If installing the LED White Light Plate, place the plate on top of the UV transilluminator and
connect the power cord coming from inside the darkroom to the back of the Plate.
3.
If installing epi UV modules, place the modules in the brackets located at the top of the
darkroom. Plug the modules into the outlets provided inside the darkroom. Place the power
switches located on the modules in the
ON
position.