Page 1 Parameters
Value and meaning
Page 2 Parameters
Value and meaning
Selection_Method
1 = Switch Only
2 = Joystick
3 = Jo Switch
4 = Joystick Only
Default: 2 Joystick
Alarm
1 = No Alarm
2 = Audible Alarm
3 = Audible Alarm + Text
Message
Default: 1, no alarm
Time_Base
1 = 250ms,
2 = 300ms,
3 = 350ms,
4 = 450ms, (Default)
5 = 600ms,
6 = 800ms,
7 = 1.2sec,
8 = 2 sec
Max_Profile_Number
Can be set from 1-5.
Default: 5
Avoid_Accidental_Hits
1 = No accidental hits are not
avoided (default)
2 = Yes, during mode
selection.
3 = Yes, during startup
4 = Yes, both
RJM
N = not fitted
Y = yes, RJM is fitted
Default: not fitted.
Switch 2
1 = Right Click
2 = Left Double Click
3 = Left Drag Lock
4 = Mouse Speed
5 = Exit mouse mode
Default: 1 Right Click
Enable_My_Set
N = No, do not use my set of
operation modes
Y = yes, use my set of
operation modes.
Default: No
Power_Down_Timer
N = not required
Y = yes, power down the
Genie after 20mins of no use.
Default: Yes
Joystick_Threshold
1 = 12.5%
2 = 25%
3 = 50%
Default: 25%
Also effects mouse movement
sensitivity (see text)
Does not apply to RJM joystick
(fixed at 50%)
Exit_Time
1 = 500ms
2 = 1 sec exit time
3 = 2 sec (default)
4 = 3 sec exit time
5 = 4 sec exit time
6 = 5 sec exit time
7 = 6 sec exit time
8 = 7 sec exit time
External_Device
1 = Standard connection.
2 = 2 cable connection to a
DynaVox communication aid.
3 = 2 cable connection to a
GEWA Progress ECU
Default: 1
To set a parameter
1.
Make sure the Genie Joystick is on.
2.
Select program mode by pressing the program switch.
3.
Page 1 is automatically selected.
4.
The number of times the decimal point flashes indicates what page you are on. To select the
next page press switch 2.
5.
Select the parameter by clicking switch 1 until the corresponding number of battery LEDs are
lit.
6.
Change the parameter by holding down switch 1 until the unit beeps and the value displayed in
the 7 segment display changes.
7.
To exit program mode step through the remaining page(s) by clicking Switch 2 until the decimal
point flashes very rapidly.
To reset the parameters to factory values
1.
Make sure the Genie Joystick is on.
2.
Press and hold the program switch for about 3 seconds until the decimal point flashes rapidly,
then release the program switch.
3.
All parameters have been reset to their factory (default) values.
Before you do any programming, look at the these tables of parameters and know beforehand what
changes you want to make. Be absolutely sure about changing a user’s parameter settings. It is a
good idea to photocopy this sheet and make note of a user’s parameter settings.
To modify the seat function list
1.
Enter program mode and select page 3.
2.
The first five parameters determine what actuator is
controlled by what seat symbol.
3.
Change a parameter in the normal way (see opposite
text). Each parameter can be set from 1 to 6. Setting a
parameter to 6 means that seat symbol will not be
used.
4.
See text for further explanation.
To create your own set of operation modes
5.
Enter program mode and select page 4.
6.
The symbol in the 7 segment display indicates the first
mode.
7.
Use short clicks of Switch 1 to select the mode you
want for this step.
8.
When it is displayed press the primary switch for a long
time, until the Genie beeps. The chosen mode for this
step is stored and a new step begins.
9.
Repeat from step 3 until you have entered all the steps
of your set. Note that you can not exceed 6 steps.
10.
When you have finished entering your set press and
release switch 2. The steps are stored in the ClickToGo
and normal operation resumes.
11.
To enable your set you must re-enter program mode
and change the parameter ‘Enable_Set’ on page 2 from
‘N’ to ‘Y’.