33
6
Hold Function
33
Hold Function
Chapter
6
- Operation of peak hold (angle) (start mode: level)
After pressing the
key, hold operation starts when torque and angle exceeds the start level
from below repetitively. When the torque falls below the stop level, detection will be stopped and
the torque will be fixed. After the torque falls below the stop level, the torque will be compared
with the start level again.
Press the
key again to stop hold operation.
Detection/hold
section
Angle
Torque
Indicated value
HOLD display
Hold fix section
key input
HOLD
Start level
Stop level
Detection/hold
section
Lights off
Flashes
Lights on
Flashes
Lights off
Angle hold value
HOLD
HOLD