Torque Settings
17
Torque Settings
Chapter
4
Decimal point (DECIMAL P.)
(These items only available in case the UTM capacity is set to OTHER)
Set the decimal point of torque to display.
<Setting range>
0000, 000.0, 00.00, 0.000
Sign (SIGN)
Set the sign of torque.
<Setting range>
NORMAL, REVERSE, ABSOLUTE
- NORMAL
Viewing the UTM
Ⅱ
from the drive side, the direction in which the load side is
twisted counterclockwise is displayed as the positive direction.
- REVERSE
Viewing the UTM
Ⅱ
from the drive side, the direction in which the load side is
twisted clockwise is displayed as the positive direction.
- ABSOLUTE
Shows the absolute value of torque without the direction of the load.
Zero calibration (ZERO CAL.)
Zero point is registered while torque load is not applied to UTM
Ⅱ
. Furthermore, if the initial torque
remains in the jig and so on for the purpose of measurement, the zero point may be registered while
the initial torque is being applied; however, the initial torque range needs to be subtracted from the
signal input range.
<Setting range>
-5.000 to 5.000 [V]
1.
Press the
key on the main screen and enter
the setting value display screen.
2.
Display the "CONFIG TORQUE" screen using the
keys.
3.
Select zero calibration
using
the
keys.
4.
Unload the sensor and press the
key.
Minimum scale (MIN. SCALE) (optional if there is no change)
Set the minimum unit of torque such as scale interval and scale division.
<Setting range>
1, 2, 5, 10, 20, 50, 100
SET
▼
▲
▲
▼