35
7
Graph
35
Gr
aph
Chapter
7
*1: The plotting time can be changed by setting the maximum recording time of setting values
as follows (thin-out):
5 seconds
No thin-out
20ksps
10 seconds
Take 1 data from 2 samples
10ksps
20 seconds
Take 1 data from 4 samples
5ksps
40 seconds
Take 1 data from 5 samples
4ksps
50 seconds
Take 1 data from 10 samples
2ksps
100 seconds
Take 1 data from 20 samples
1ksps
If sample hold, peak hold or bottom hold is operated during graph plotting, the graph will be
marked.
Plot mode (PLOT MODE)
Sets the repeat condition for plotting operation.
<Setting range>
REPEAT, SINGLE
- REPEAT
Once the
key is pressed, plotting is repeated until the
key is
pressed again.
- SINGLE
Plots once if the
key is pressed once.
Start mode (START MODE)
The plotting start operation changes according to the start mode.
<Setting range>
KEY, EDGE
↑
, EDGE
↓
- KEY
Starts taking data and plotting a graph if the
key is pressed.
- EDGE
↑
Starts taking data and plotting a graph when the torque crosses the level from
below to above after the
key is pressed.
* The level is set by the setting value "start level".
- EDGE
↓
Starts taking data and plotting a graph when the torque crosses the level from
above to below after the
key is pressed.
* The level is set by the setting value "start level".
Start level (START LEV.)
Sets the level to be crossed.
* Enabled only when the start mode is EDGE
↑
or EDGE
↓
.
<Setting range>
-9999 to 9999
SAVE
SAVE
SAVE
SAVE
SAVE
SAVE