31
6
Hold Function
31
Hold Function
Chapter
6
6-6.
Average hold
Holds the average value of torque.
- Operation of average hold (start mode: key)
Holds the average value after pressing the
key as the indicated value.
Press the
key again to confirm.
Press the
key again to stop hold operation.
- Operation of average hold (start mode: level)
The hold operation starts when the torque passes the start level (from above to below or from below
to above) after pressing the
key. Afterwards, the hold operation continues until the torque
passes the stop level (in the opposite direction of the start level). After the passing the stop level, the
hold data is displayed while monitoring the indicated value to pass the start level.
Press the
key again to stop hold operation.
Time
Indicated value = 0
Torque
Indicated value
HOLD display
key input
HOLD
Hold fix
Detection/hold section
Lights off
section
Hold release
section
Flashes
Lights on
Lights off
HOLD
HOLD
HOLD
Time
Torque
Indicated value
Detection/hold
HOLD display
section
Hold fix section
key input
HOLD
Start level
Stop level
Detection/hold
section
Lights off
Indicated value = 0
Hold release
section
Flashes
Lights on
Flashes
Lights
on
Lights off
HOLD
HOLD