18.Interface
140
18-4-4. Writing Sampled Waveform Data
(00 ~ 63)
W 5 2 SP 0
Host
Terminator
32 pieces of data
Offset
1: High Limit
0: Low Limit
Offset binary
4-digit hexadecimal number
Check sum
・
In the Multi-Hold mode and Hysteresis mode, the operation is
executable only when the set value of W51 is 0 ~ 15 or 92.
・
In the Waveform Comparison mode and Waveform and Displacement
Comparison mode, the operation is executable only when the set
value of W51 is 0 ~ 15, 91 or 92.
W 5 2 SP 0
F395
W 5 2 SP 1
F395
Terminator
Terminator
Return when the check sum is correct
1: High Limit Offset
(00 ~ 63)
(Sent)
Return when the check sum is incorrect
0: Low Limit
1: High Limit Offset
(00 ~ 63)
(Sent)
0: Low Limit