14.Waveform Editing
82
14. Waveform Editing
14-1. Waveform Call-up
Stored waveforms are called up for editing by this operation. Waveforms are stored
under "14-4-4. Waveform Save" P91, and edited under "14-4. Waveform Editing" P84.
Setting range: 0 ~ 15
After call-up, go to the waveform editing screen to display the called-up waveform.
14-2. Waveform Clear
Waveform data for waveform sampling that have been temporarily registered inside is
cleared by this operation. Before sampling waveforms, be sure to execute this operation.
MODE
→
Wave Sampling
→
Wave Call
No pass/fail judgment can be made on the waveform as it is called up
here. For making a judgment, the waveform should be stored under
"14-4-4. Waveform Save" P91, and further the ch should be input
externally (See "13-3. Waveform Sampling Procedures" P69.)
MODE
→
Wave Sampling
→
Clear
High limit waveform: Set to -9999 (Not including the decimal point.)
Low limit waveform: Set to 9999 (Not including the decimal point.)
The operation is executed by pressing the
key.
The waveform data for comparison from channel 0 to 15 that
have been registered inside is not cleared.