8.8 Simulation commands
Table 8.42 –
continued from previous page
Column
Example value
Description
24-33
00244
Simulated Temperature ADC value. See Equation 8.1 .
34
x
Terminating character.
An example command is:
S,0000000360,0000000360,0000000360x
0123456789 123456789 123456789 1234
Table 8.43: Response of setting simulation values (S...x)
Column
Example value
Description
0-1
S,
Confirmation of S...x command.
2-13
0000094000c,
Simulated counts.
14-25
0000000000f,
Simulated frequency in Hz.
26-37
0000000245t,
Simulated temperature ADC value. See Equation 8.1 .
38-39
r,
Beginning of calculated readings.
40-47
18.04m,
Calculated
mag
arcsec
2
.
48-60
0000000000Hz,
Frequency used for calculation.
61-72
0000094000c,
Counts used for calculation.
73-85
0000000.204s,
Calculated period from counts.
86-92
029.0C
Temperature used for calculation.
93-94
Carriage return (0x0d), Line feed (0x0a).
An example response is:
S,0000094000c,0000000000f,0000000245t,r, 18.04m,0000000000Hz,0000094000c,0000000.204s, 029.0C
0123456789 123456789 123456789 123456789 123456789 123456789 123456789 123456789 123456789 1234
Unihedron SQM-LU-DL Operator’s Manual -
20220119
89
Summary of Contents for SQM-LU-DL
Page 2: ......
Page 17: ...17...
Page 19: ...Figure 4 1 Front and back of of unit Unihedron SQM LU DL Operator s Manual 20220119 19...