8 Commands and responses
Table 8.25 –
continued from previous page
Column
Example value
Description
20-23
xxxx
(Optional snow accessory)
Snow reading; Big endian raw 32 bit binary value for magnitudes.
The decimal reading value is determined as:
Reading
M P SAS
=
xxxx
6553600
.
0
.
24-27
xxxx
(Optional snow accessory)
Snow frequency, Big endian 32 bit value.
28-31
FFFFFFFF
hex
Spare four bytes.