8.7 Setting Interval reporting parameters
8.7.3
Interval setting response
Either making the request “Ix” (note upper case “I”) or any request to set the interval report setting produces the
following response shown in table 8.40:
Table 8.40: Response of viewing or setting interval reporting parameters
Column
Example value
Description
0
I
Interval settings from EEPROM and RAM are being returned.
2-12
0000000360s
Interval period that was set into EEPROM.
14-24
0000300360s
Interval period that was set into RAM.
26-37
00000017.60m
Threshold value that was set into EEPROM.
39-50
00000017.60m
Threshold value that was set into RAM.
51-52
Carriage return (0x0d), Line feed (0x0a).
An example response is:
I,0000000360s,0000000360s,00000017.60m,00000017.60m