8 Commands and responses
8.6 Continuous reporting commands
The Continuous reporting features are somewhat experimental. A description of each feature follows:
Table 8.36: Summary of standard commands
Option
Description
Reporting enabled
Send a reading result every time it is available from internal computations. In
bright settings, a reading will be made available once per second, around 14mpsas,
a reading will be available depending on the sensor response which is between 60+
times per second and 60 seconds in very dark settings.
Ideal crossover firmware
Changes the crossover of reporting from original (around 15mpsas to 12.32mpsas
which results in better resolution across the entire range of possible readings.
Reporting compressed
Send the above enabled reading out in a compressed format without extra text.
Report un-averaged
Sends the above reading out from the un-averaged accumulator.
Normally the
averaged reading is sent.
Table 8.37: Continuous reporting command summary
Command
Description
Yx
Get status of the continuous reporting features.
YRx
Enable continuous reporting.
Yrx
Disable continuous reporting.
YCx
Enable new crossover firmware code.
Ycx
Disable new crossover firmware code.
YPx
Enable continuous reporting compression.
Ypx
Disable continuous reporting compression.
YUx
Enable un-averaged continuous reporting.
Yux
Disable un-averaged continuous reporting.
Table 8.38: Response of all continuous reporting requests
Column
Example value
Description
0
Y
Confirmation of command.
1
r
R = Continuous reporting enabled.
r = Continuous reporting disabled.
2
C
C = Ideal crossover firmware code enabled.
c = Ideal crossover firmware code disabled.
3
p
P = Continuous reporting compression enabled.
p = Continuous reporting compression disabled.
4
u
U = Continuous reporting un-averaged readings enabled.
u = Continuous reporting un-averaged readings disabled.