7 Unihedron Device Manager
An override from the command line options (-LCTH,x) can be made. The override will only take effect for the current
session and will not alter the saved configuration.
Pre-reading audio alert
In cases where you want to be alerted that a reading will be taken, for example, when manually
pointing the meter, the Pre-reading audio alert (Fig. 7.29) can be enabled to play an audio file two seconds before the
record is logged. The Test button plays this file for test purposes. The audio file is named
prereading.wav
and is located
in the Data Directory (which can be identified from the main page
View
→
Directories
).
Figure 7.29: Log Continuously Pre-audio alert
Alarm for darkness
When the darkness is over the threshold amount, an audio file can be enabled to play (Fig. 7.30) .
The Test button plays this file for test purposes. The audio file is named
alarmsound.wav
and is located in the the Data
Directory (which can be identified from the main page
View
→
Directories
).
Figure 7.30: Log Continuously Alarm for darkness
Transfer
The .dat file can be sent via FTP or SCP to a server by by making settings on the left side of the Transfer tab
(shown in Fig. 7.31)
Figure 7.31: Log Continuously Transfer tab
SCP transfers require that the
scp
command exists externally from UDM. For Linux and Mac, the
scp
command is
scp
and can be proven to exist with the
which scp
command. In Windows, you may have to install a program called
pscp
and rename it to
scp
.
Annotation settings
While the “Log Continuous” mode is recording data, annotations can be made to the log file records
with hotkeys. The Annotation tab allows the definition of hotkeys and their associated annotation text that will appear
at the end of an annotated record.
46
Unihedron SQM-LU-DL Operator’s Manual -
20220119