User’s Manual UT5583 Insulation Resistance Tester
Instruments.uni-trend.com
36
/
47
default is falling edge. Low level maintains 1ms.
After receive this signal, the instrument will perform a complete
period test (Charge --> Test --> Discharge --> Stop).
10
STOP
Stop test
Pulse signal, low level maintains 10~50 ms (typical value is 30 ms.
)
11
KLOCK
Lock Key
Pulse signal, low level maintains 10~50 ms (typical value is 30 ms.
)
This signal is only for lock, unlock should operate on the instrument.
Table 12-5 Pin Definition of Power Terminal in Period Comparison Mode
Pin
Name
Description
19
ISO-COM
Common ground, not allow to float.
Be sure to be reliably connected to the ground (COM) port of the power
supply of an external controller (such as a PLC).
20
21
24
Internal
ISO-VCC
Output
Positive output of internal VCC power supply
Internal isolation power voltage: 5V, 0.2A, 1Wmax
Do not connect if it is not necessary.
25
1. Pin P24/P25 is output terminal of internal isolation power.
2. Warning: Set pin P24/P25 to be floating when normal connecting to PLC.
3. Internal power voltage is limited (5V
,
0.2A, 1Wmax), it cannot drive the power
relay or high power LED.
4. Pin P24/P25 has built-in automatic self-recovery fuse, it will cut off the
connecting of ISO-VCC when it excceds 0.5 A, long short circuit will cause the
fuse to burn out.
12.3.2 Time Sequence of Period Comparison Mode
Figure 12-5 Time Squence of Sampling Data during the Test State in Period Comparison Mode, BIN is
Comparsion Output
Table 12-6 Time Squence of Period Comparison Mode
Description
Typical Value
T1
Trigger pulse width
1ms
T2
Measurement period
Trigger delay
Relate to Setup, refer to section 7-
Trigger Delay