User’s Manual UT5583 Insulation Resistance Tester
Instruments.uni-trend.com
21
/
47
6.2
Measurement Result
6.2.1
Display Page of Measurement Result
Figure 6-2 Status Icon (Test Page)
Table 6-1 Descriptio of Measurement Result
No.
Description
1
Real-time monitoring output voltage
2
Timer of charge, test or discharge
3
State display
Display the dicharge state during stop test or discharge timing.
4
Total comparative results
Qualified result displays tick with green, unquailied result displays in red fork.
5
Comparator state display
Qualified: PASS
Over the upper limit: UFAIL
Over the lower limit: LFAIL
Open circuit (poor contact): OPEN
6
Insulation resistance display
It displays in white character during the test and in gray character during
discharge/stop state.
Note: The instrument will direct display measuring range for each scale.
As shown in Figure 6-2, measuring range displays [1] Auto [40.0 kΩ,420 kΩ], it
represents the measuring range number is 1, the measuring range mode is automatic
and the measuring range of resistance test is 40.0 kΩ~420 kΩ.
If range displays [1] Lock [40.0 kΩ, 420 kΩ], the tested resistance is lower than 40.0
kΩ, it displays Under.F, which means the range is over the lower limit; the tested
resistance is higher than 420 kΩ, it displays Over.F, which means the range is over the
upper limit.
7
Leakage current display
It only display when display mode sets to resistance/current.