User’s Manual UT5583 Insulation Resistance Tester
Instruments.uni-trend.com
28
/
47
Baud Rate
9600, 19200,
38400, 57600,
115200
115200
Baud rate of serial bus
Station
Number
0~32
01
Modbus (RTU) requires the station address
①
Instrument allow to use station number 00 to
perform broadcast communication.
②
1~32: the address of the instrument when
connect to the bus.
Result
Send
FETCH? ,
automatic
Automatic
This function is only for SCPI.
The instrument supports automatically send data to
the host computer. The data will be automatically
sent to the host computer after each test is
completed
without the need for the host to send the
FETCH? command.
IP Address
192.168.030.036
It can set by numeric keyboard, it will be used when
select LAN interface.
Port
Number
502
It is default setting and not easy to change. It will be
used when select TCP interface.
Power
Frequency
50 Hz, 60 Hz
50 Hz
Power frequency is selected based on the power
supply.
Factory
Setting
Restore
All settings of the instrument will restore to the
factory setting.
<File Mangement> page presets to file 1.
9.2 <System Information> Page
Enter <System Configuration> page and press function key to select [System Information].
This part inculdes model’name, serial number and version of the instrument.
This page does not require user involvement.
Figure 9-2 <System Information> Page