12.3 Handler Interface — Period Comparison Mode
Table 12-5 Period Comparison Mode
In addition, period comparison mode requires normal output the sorting results, so [Test Time] in
<Setup> page should set to the sepecified time
12.3.1 Wiring Signal of Period Comparison Mode
Table 12-3 Pin Definition of Output Terminal in Period Comparison Mode
Pin
Name
Description
17
EOM
0: read 1: wait
In test page, it is high level valid when sampling, and turn to low leven
when sampling is finished.
14
PASS
Qualified output of comparator
0: PASS (Low level is valid.)
15
FAIL
Total disqualified output of comparator
0: FAIL (Low level is valid.)
1
HI
Over the upper limit disqualified output of comparator
0: HI (Low level is valid.)
2
LO
Over the lower limit disqualified output of comparator
0: LO (Low level is valid.)
16
OPEN
Poor contact (open circuit)
0: OPEN (Low level is valid.)
In opern circuit, all sorting signals (PASS, FAIL, HI LO) are restore to
high level.
Table 12-4 Pin Definition of Intput Terminal in Period Comparison Mode
Pin
Name
Description
13
TRIG
Test trigger input terminal
Edge signal, trigger edge tyoe can select in <Setup> page, the