5. Inspection and Installation
5.1 Packing List
Before using the instrument,
1. Check the appearance whether is damaged or scratched;
2. Check the packing list if has loss.
If the product is damaged or accessory is missing, please contact UNI-T sales department or
distributor.
Article
Quantity
Remarks
Insulation
resistance tester
1 piece
Power cord
1 piece
RS232
communication
line
1 piece
Test line
1 set
1 piece of red, green and black
Spare fuse
2 pieces
T1AL for 220V
Or T2AL for 110V
Pluggable wiring
terminal
1 piece
Footswitch
Download manual
1 piece
User’s manual
0 piece
Electronic user’s manual can download
from the official website.
5.2 Power Requirements
The instrument is deign to use under CAT
Ⅱ
, do not use under CAT
Ⅲ
and CAT
Ⅳ
.
Befor boot the power, please make sure the power voltage and fuse is consistent with the voltage
selected by the AC SELECTOR switch on the back panel of the instrument.
Input Voltage
Frequency Range
Fuse
(slow-blow)
Rated
Power
110V
47-63Hz
2A
30VA
220V
1A
Warning: To prevent electric shock, please make sure that the power line is securely connect to the
ground.
5.3 Operation Environment
UT5583 insulation resistance tester is recommended for use under the following environmental
conditions.
Environmental Requirements