Page
62
Ulthera, Inc
H
Code H
Transducer motion not detected
Please remov e and reinsert the
transducer.
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
Transducer motion was not detected.
Ensure that the transducer is properly
mounted in the handpiece. Please be
sure to always hit Scan N before
remov ing transducer. Remov e and
reinsert the transducer.
I
Code I
Communication halted
Please restart the system.
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
Communication halted due to an
ev ent detected in the control unit.
J
Code J
Handpiece communication
halted
Please restart the system.
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
Communication halted due to an
ev ent detected in the control unit.
K
Code K
Software halted
Please restart the system.
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
Software was halted due to an ev ent
detected in the control unit.
L
Code L
Transducer out of lines
Please replace transducer and
continue.
See User’s Manual for further
information.
The Transducer’s remaining line count is
zero. Remov e and replace the
Transducer.
M
Code M
Handpiece motion halted
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
Inspect handpiece. Ensure that the
transducer is properly mounted and
latched in the handpiece.
Summary of Contents for DeepSEE DS 4 - 4.5
Page 2: ...Page 2 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 38: ...Page 38 Ulthera Inc 6 1 1 Face Neck Amplify...
Page 39: ...Ulthera Inc Page 39 6 1 2 General Regions Face Neck...
Page 69: ...Ulthera Inc Page 69 THIS PAGE INTENTIONALLY LEFTBLANK...
Page 70: ...Page 70 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 71: ...Ulthera Inc Page 71 THIS PAGE INTENTIONALLY LEFTBLANK...