Page
16
Ulthera, Inc.
Figure 3.3
Handpiece with transducer inserted, top and side v iews.
Table 3.2
Handpiece and Transducer Description
ITEM
DESCRIPTION
1
SEE
Pushbutton
•
Engages IMAGING state (if not already imaging)
•
Places system in READY state (times out in 40 seconds)
•
Stops TREATING if treatment in progress
2
TREAT
Pushbuttons
Engages TREATING state
3
Latch
Locks transducer into handpiece
4
Transducer
Image/treat transducer
5
Strain Relief / Cable Connects handpiece to Control Unit
3.2.3.
Transducers
Figure 3.4 is an illustration of an image/treat transducer. The transducer can image and treat a
region of tissue up to 25 mm long and can image a depth of up to 8 millimeters. Treatment
occurs along a line less than or equal to the transducer’s active length, which is indicated by
guides on the sides of the transducer, as described in
Table 3.3. An additional guide at the front tip of the transducer represents the
center of the treatment line. In therapy mode, bursts of sound energy create
a linear sequence of individual, discrete, thermal coagulation points (TCPs).
A label atop the transducer provides the transducer type, expiration date,
and other information.
Figure 3.4
Image/Treat Transducer, separated from handpiece (see
Summary of Contents for DeepSEE DS 4 - 4.5
Page 2: ...Page 2 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 38: ...Page 38 Ulthera Inc 6 1 1 Face Neck Amplify...
Page 39: ...Ulthera Inc Page 39 6 1 2 General Regions Face Neck...
Page 69: ...Ulthera Inc Page 69 THIS PAGE INTENTIONALLY LEFTBLANK...
Page 70: ...Page 70 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 71: ...Ulthera Inc Page 71 THIS PAGE INTENTIONALLY LEFTBLANK...