7.3.3.
Select a transducer:
The Ulthera System has six types of transducers:
Table 7.5
Types of Transducers
TRANSDUCER
SPECIFICATIONS
DS 10 – 1.5
Low energy lev el and 1.5 mm focal depth.
DS 10 – 1.5N
Low energy lev el, 1.5 mm focal depth, and a narrower contact area
than the DS 10 – 1.5.
DS 7 – 3.0
Low energy lev el and 3.0 mm focal depth.
DS 7 – 3.0N
Low energy lev el, 3.0 mm focal depth, and a narrower contact area
than the DS 7 – 3.0.
DS 7 – 4.5
Intermediate energy lev el and 4.5 mm focal depth.
DS 4 – 4.5
High energy lev el and 4.5 mm focal depth.
1.
Check the expiration date on the transducer package.
2.
Open the sealed pouch.
3.
Connect the transducer to the handpiece by sliding the transducer into
the handpiece until the passive latch locks into place.
Figure 7.7
Connecting the Transducer
A tone will sound when the transducer has been correctly inserted. The
Treatment Guideline area on the
DeepSEE
screen will display the regions
available for treatment with the inserted transducer.
Caution: If a warning or caution message is displayed, or a message
that says “Transducer Not Connected,” disconnect and reconnect the
transducer. If the problem persists, contact your Ulthera representative.
Caution: If the handpiece or transducer is dropped or broken, or any
part of the system is damaged, disconnect the system from the power
source outlet before touching any other part of the equipment.
Thoroughly inspect the equipment for external damage before
reconnecting. Do not use a damaged handpiece or transducer.
Summary of Contents for DeepSEE DS 4 - 4.5
Page 2: ...Page 2 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 38: ...Page 38 Ulthera Inc 6 1 1 Face Neck Amplify...
Page 39: ...Ulthera Inc Page 39 6 1 2 General Regions Face Neck...
Page 69: ...Ulthera Inc Page 69 THIS PAGE INTENTIONALLY LEFTBLANK...
Page 70: ...Page 70 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 71: ...Ulthera Inc Page 71 THIS PAGE INTENTIONALLY LEFTBLANK...