Treatment Guidelines
6.1.
Preset Guidelines and Energy Levels
The Ulthera System is programmed with preset guidelines that have been
established through clinical experience. Table 6.1 describes the preset
guidelines available on your System.
Table 6.1
Guideline Names and Energy Lev els
Guideline Name
Energy Level
Range
Default
Energy Level
Face & Neck (Amplify)
1 – 4
2
Chest
DS 7-3.0 & DS 4-4.5
1 – 4
4
DS 10-1.5
1 – 4
3
General Regions (Face & Neck)
1 – 4
2
Train: Face & Neck (Amplify)
0 only
0
Train: Chest
0 only
0
To facilitate training with your system, specific training guidelines have been
pre-programmed into the guideline list as well. These are noted as “Train: …”
These training guidelines should be used for training purposes only where no
energy delivery is desired. Lines available for the transducer will not be
decremented while in a training guideline.
If regions other than those depicted in the preset guidelines are desired, it is
recommended that you utilize the
User Regions
function as described in
Section 7.2.3. This function will allow you to define text-based regions to select
and treat against.
Each transducer is programmed with set energy levels. Table 6.2 describes
the energy levels available for each transducer.
Summary of Contents for DeepSEE DS 4 - 4.5
Page 2: ...Page 2 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 38: ...Page 38 Ulthera Inc 6 1 1 Face Neck Amplify...
Page 39: ...Ulthera Inc Page 39 6 1 2 General Regions Face Neck...
Page 69: ...Ulthera Inc Page 69 THIS PAGE INTENTIONALLY LEFTBLANK...
Page 70: ...Page 70 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 71: ...Ulthera Inc Page 71 THIS PAGE INTENTIONALLY LEFTBLANK...