Warning: The Ulthera system should not be used adjacent to, or stacked
with other electronic equipment. If the system must be installed in close
proximity to other equipment, both Ulthera and the nearby equipment
should be observed to verify normal operation in that configuration.
4.5.2.
Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
Table 4.6
Electromagnetic Immunity
IMMUNITY TEST
IEC 60601 TEST
LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC
ENVIRONMENT -
GUIDANCE
Electrostatic
discharge
(ESD)
IEC 6100-4-2
±6 kV contact
±8kV air
±6 kV contact
±8kV air
Floors should be wood,
concrete or ceramic
tile. If floors are cov ered
with synthetic material,
the relativ e humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for
input/output lines
± 2kV for power
supply lines
±1kV for
input/output lines
Mains power quality
should be that of a
typical commercial or
hospital env ironment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality
should be that of a
typical commercial or
hospital env ironment.
Voltage dips,
short
interruptions
and v oltage
v ariations on
power supply
input lines IEC
61000-4-11
<5%
U r
(>95% dip in
U r )
for 0,5 cycle
40%
U r
(60% dip in
U r )
for 5 Cycles
70%
U r
(30% dip
in
U r )
for 25
cycles
<5%
U r
(>95% dip
in
U r )
for 5 sec
<5%
Ur
(>95% dip in
Ur
)
for 0,5 cycle
40%
U r
(60% dip
in
U r )
for 5 cycles
70%
Ur
(30% dip in
Ur
) for 25 cycles
<5%
Ur
(>95% dip
in
Ur
) for 5 sec
Mains power quality
should be that of a
typical commercial or
hospital env ironment. If
the user of the Ulthera
system requires
continued operation
during power mains
interruptions, it is
recommended that
Ulthera be powered
from an uninterruptible
power supply.
Power
Frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m at 50 and
60Hz
Power frequency
magnetic fields should
be at lev els
characteristic of a
typical location in a
typical commercial or
hospital env ironment.
Summary of Contents for DeepSEE DS 4 - 4.5
Page 2: ...Page 2 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 38: ...Page 38 Ulthera Inc 6 1 1 Face Neck Amplify...
Page 39: ...Ulthera Inc Page 39 6 1 2 General Regions Face Neck...
Page 69: ...Ulthera Inc Page 69 THIS PAGE INTENTIONALLY LEFTBLANK...
Page 70: ...Page 70 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 71: ...Ulthera Inc Page 71 THIS PAGE INTENTIONALLY LEFTBLANK...