Chapter 4 Flow measurement
- 34 -
W-method Installation
W-method is usually used on plastic pipes with a diameter from 10 millimeters to
100 millimeters
N-method Installation
Rarely used method.
4.3.1.1.
Input measurement parameters to get the installation distance
Before starting the measurement, the flowmeter needs to be initially set, usually
from the menu 10 to 26 item by item (M39 menu is available in a variety of
languages). After setting, the installation distance of the sensor can be obtained in
M25, which refers to the distance between the innermost edge of the two sensors
(see figure above), and the sensor can be installed according to this data.
4.3.1.2.
Handling installation point
There are two installation points of the external clamp sensor, which are the
upstream sensor and the downstream sensor.In dealing with these two Ann
When decorating, the treatment area of one installation point is about the same
size as the sensor, and the treatment area of another installation point is about the
same size.It should be about twice the size of the sensor area (centered on the
installation point) to facilitate signal debugging.First clean the area outside the tube
where the sensor is to be installed to remove rust and paint.If there is antirust layer
should also be removed, it is best to use a grinding machine to polish the metallic
luster, and then wipe away oil and dust with a clean cloth.
4.3.1.3.
Sensor fixation
Apply 2~3mm thick random coupled agent on the signal generator of the sensor,
Summary of Contents for MC-SUP2000H
Page 1: ...Medidor de flujo ultrasónico en líquidos portátil Meacon MC SUP2000H ...
Page 5: ......
Page 11: ...Chapter 1 Introduction 6 1 6 Parts Identification 1 6 1 Converter Front view Side view ...
Page 16: ...Chapter 1 Introduction 11 View and size Small size HS Medium size HM ...
Page 17: ...Chapter 1 Introduction 12 High temp Small size HS HT ...
Page 18: ...Chapter 1 Introduction 13 High temp Medium size HM HT Extension bracket ...
Page 69: ......