Chapter 7 Troubleshooting
- 56 -
7.3. Causes of abnormal measured values and treatment methods
Status
Reason
Handle
The measured value
shows a negative value
display
The link between the host
and the sensor (upstream
and downstream sensors)
is reversed.
Make the correct
connection.
The actual flow is reverse.
The measured value
drifts abnormally when
the flow rate is constant
The straight pipe is not
long enough.
Move to a position where
the length can be ensured
(upstream 10D,
downstream 5D).
There are pumps, valves,
etc. nearby that cause fluid
flow turbulence.
Keep the installation
distance above 30D.
There is actually a
pulsation.
Through the damping
setting, increase the
corresponding time.
Ultrasonic waves cannot
propagate into the
pipeline, and the
measured value remains
unchanged.
1. Improper setting
After confirming the reason,
temporarily remove the
sensor, clean the
installation part and apply
couplant, and finally install
the sensor in a place slightly
staggered from the previous
part.
*The pipe specification is
wrong.
*A sensor is installed at the
welding part.
*The sensor size is wrong.
*Insufficient application of
Summary of Contents for MC-SUP2000H
Page 1: ...Medidor de flujo ultrasónico en líquidos portátil Meacon MC SUP2000H ...
Page 5: ......
Page 11: ...Chapter 1 Introduction 6 1 6 Parts Identification 1 6 1 Converter Front view Side view ...
Page 16: ...Chapter 1 Introduction 11 View and size Small size HS Medium size HM ...
Page 17: ...Chapter 1 Introduction 12 High temp Small size HS HT ...
Page 18: ...Chapter 1 Introduction 13 High temp Medium size HM HT Extension bracket ...
Page 69: ......