Chapter 4 Flow measurement
- 30 -
4.1.3. Scaling
Scaling on the inner wall of the pipe will attenuate the transmission of ultrasonic
signal and reduce the pipe inner diameter.Therefore, the scaling of the inner wall of
the pipe will make the flow meter can not be measured normally or affect the
measurement accuracy.Therefore, it is necessary to avoid choosing the place of
scaling on the inner wall of the pipeline as the installation point.If the scaling
installation point cannot be avoided, the following measures can be taken to
Summary of Contents for MC-SUP2000H