48
Operative Instructions
12.6 Alarm table
ALARM
DESCRIPTION
ACTION
E00 EEPROM ALARM
Firmware installed successfully.
1- Press RESET
2- Install new dataset
E01 BLACKOUT
A blackout occurred during cycle
execution and the program was
stopped.
Press RESET
E02 DOOR OPEN
The door is open or unlocked.
Close the door.
Press RESET
If it persists, call the service personnel.
E07 NO LOCKED DOOR
The door did not lock within the set
time.
Press RESET
If it persists, call the service personnel.
E09 NO UNLOCKED DOOR The door did not unlock within the
set time.
Press RESET
If it persists, call the service personnel.
E11 DIRTY WATER
Check for dirt on the load and check
the inlet demineralized water.
Press RESET
If it persists, call the service personnel.
E20 NO COLD WATER
The cold water tap could be closed or
partially open. Check it.
Check if the safety floatswitch has not
tripped.
Press RESET
If it persists, call the service personnel.
E21 NO HOT WATER
The hot water tap could be closed or
partially open. Check it.
Press RESET
If it persists, call the service personnel.
E22 NO DEMI WATER
The demineralized water tap could
be closed or partially open. Check it.
Press RESET
If it persists, call the service personnel.
E26 PRINTER ERROR
The printer is turned off, unplugged
or out of paper. Check it.
Press RESET
If it persists, call the service personnel.
E30 FLOWMETER 1
Chemical liquid tank 1 might be
empty.
Check it.
Press RESET
If it persists, call the service personnel.
E31 FLOWMETER 2
Chemical liquid tank 2 might be
empty.
Check it.
Press RESET
If it persists, call the service personnel.
E32 FLOWMETER 3
Chemical liquid tank 3 might be
empty.
Check it.
Press RESET
If it persists, call the service personnel.
E33 FLOWMETER 4
Chemical liquid tank 4 might be
empty.
Check it.
Press RESET
If it persists, call the service personnel.
E34 PRODUCT TIMEOUT
The density of the chemical liquid
might be too high. Check it.
Press RESET
If it persists, call the service personnel.
E41 DRAIN TIMEOUT
The drain pipe might be blocked or
clogged. Check it.
Press RESET
If it persists, call the service personnel.
Summary of Contents for TIVA8-MD
Page 56: ...56 Operative Instructions Notes...
Page 57: ...57 Operative Instructions Notes...
Page 58: ...58 Operative Instructions Notes...
Page 59: ...59 Operative Instructions...