41
Operative Instructions
ID Parameter Name
Min
Max
Description
Factory
Default
14 Program button 2
1
40
Selects the program to be combined
with button 2
2
15 Program button 3
1
40
Selects the program to be combined
with button 3
3
16 Printer language
1
8
Selects the desired language for the
printer
ENGLISH
17 Drying enable
0=OFF
1=ON
Enables or disables the drying stage
(not avalable on models without drying
system)
ON
18 Lower Spray Arm
0=OFF
1=ON
Enables or disables lower spray arms
OFF
19 White LED always
ON
0=OFF
1=ON
Enables or disables the washing
chamber's LED light
OFF
20 Fahrenheit
temperature
0=OFF
1=ON
Sets the temperatures display in °F
degrees
OFF
21 PIN expiration
month
1
99
Set password expiration time
6
22 User PIN enable at
the end of cycle
0=OFF
1=ON
Available only with "User PIN
Management" option. Enables or
disables user PIN at the end of cycle
OFF
23 Upper Spray Arm
0=OFF
1=ON
Enables or disables uper spray arms
OFF
TEMPERATURES
1
Fan activation
0°C
32°F
100°C
212°F
Cooling fan operating temperature, when
the machine reaches and exceeds the set
temperature, the cooling fan turns on
50°C
122°F
TIMES
1
Time LED lights
0
999
Washing chamber light switch-on time
5 minutes
2
Fan time
0
99
Fan operating time at the end of the
cycle when the temperature is higher
than the one set
5 minutes
PUMPS
1
Cold water
flowmeter
0=OFF
1=ON
Enables or disables the cold water
flowmeter
ON
2
Cold filling time
0
1000
Sets the timer for the pump in the
absence of the flowmeter count
52
seconds
3
Hot water
flowmeter
0=OFF
1=ON
Enables or disables the hot water
flowmeter
OFF