13
Operative Instructions
- Handle the machine with appropriate trolley. Lock the
machine on the trolley with a sturdy belt (Fig. 5.3)
Do not lift the machine by grasping it from the
protruding points, such as the control panel. They
may be damaged or detached. With some metal
components there is a danger of injury or cuts.
a
WARNING! Wear cut-resistant protective
gloves during manual transport and
positioning of the machine.
Fig. 6
A label is placed outside the packaging (Fig. 6) that
indicates:
A) Minimum and maximum temperature of the
storage environment;
B) Minimum and maximum humidity of the storage
environment;
C) Minimum and maximum pressure of the storage
environment;
D) “High” position indicators;
E) Glass indicating “Fragile” material;
F) Keep the packaging dry;
G) A maximum of one machine can be stacked;
H) Do not overturn the packaging with the machine
inside;
I) It is strictly forbidden to use hooks for handling the
packaged machine.
Inside the packaging, in addition to the machine and
the documentations, there are:
- the two spray arms that will be applied on the
chamber;
- 2 drain pipe;
- the emergency release key;
- the USB key.
If the machine is to be moved, it is advisable to keep the
packaging for any other relocation.
If this is not provided, dispose of the packaging
materials: cardboard, polystyrene and other, separated
by single material, sending them to the most
appropriate final destination, which may be recovery or
storage in landfills.
4.1 Storage
If the packaged machine is temporarily stored, make
sure that it is not subjected to blows and tampering.
However, it must be placed in a closed, dry, dust-free
environment and protected from atmospheric agents.
When restored, a careful preliminary examination of its
integrity by specialised personnel is necessary.
Absolutely avoid overlapping weights or foreign bodies
on the packaging and the machine.
5 Installation
It is advisable that only furniture for professional use
be positioned in the area around the machine, to avoid
ruining them due to possible leakage of condensation
water.
Ö
IMPORTANT NOTE! For a safe installation,
the electrical disconnector/plug of the device
must be positioned free from any obstacle
and in a visible and accessible position for
the operator, so that it's easy to control
it in case of emergency or prolonged safe
disconnection.
b
HAZARD! Do not install and/or use the
washer disinfector in environments with
flammable/explosive atmosphere.
a
WARNING! Make sure that the floor is fit to
support the load of the equipment when in
operation.
a
WARNING! Make sure that the machine is
perfectly vertical and stable, using a spirit
level, if necessary.
Unevenness of the machine surface and height can be
adjusted with the two adjustable feet placed under the
machine itself.
b
HAZARD! All electrical and water connections
(loading/unloading) can only be carried out
by specialised and authorised personnel and
by consulting the relative diagrams.
Before machine positioning make sure that:
- All the components needed for installation and proper
use of the machine were installed: main switch, water
Summary of Contents for TIVA8-MD
Page 56: ...56 Operative Instructions Notes...
Page 57: ...57 Operative Instructions Notes...
Page 58: ...58 Operative Instructions Notes...
Page 59: ...59 Operative Instructions...