18
Operative Instructions
6) Thoroughly clean the edge of the salt tank.
7) Tighten the cap again.
8) Run a «Prewash» program to clean the salt residues
inside the wash chamber.
- It is absolutely necessary to remove any salt residues
that may have settled in the chamber after refill and
before rinsing.
- After each top-up, start the «Prewash» program. In
this way, any salt grains will dilute and rinse out. Salt
residues and the overflowing water and salt solution
can cause corrosion if they are not rinsed.
Following the instructions in the chapter "Starting the
Machine", start a «Prewash» program.
Ö
IMPORTANT NOTE! After the salt refill action,
the first wash cycles may fail.
Ö
IMPORTANT NOTE! The "Refill Salt Tank"
alarm may persist for few hours after the salt
refill action, until the brine is being created.
This will not affect the functionality of the
device.
5.3.5 Automatic regeneration
It is possible to have a water softening device that can
completely regenerate at precise intervals. This process
is fully automatic. Regeneration will be implemented
before the selected program is activated.
This device must be preset by the service personnel
during installation.
Regeneration can also be performed manually,
regardless of the warning message on the display.
5.4 Connection of the drain trap
A
B
1
1
3
2
2
Fig. 10
Water and drain system (Fig. 10):
A) Floor drain connection.
B) Wall drain connection.
1) Hose connection waste water.
2) Hose connection capacitor discharge machine.
3) Minimum connection point 3.9" (100 mm).
Maximum connection point 23.6" (600 mm).
The machine is equipped with a built-in drain trap
which must be mandatorily connected to the drain
system of the building.
The user must carry out periodic maintenance of the
drains and check that they are not clogged.
Before connection:
- supply the water connections with separate shut off
valves;
- flush out the water pipes that are to be connected to
the machine, to prevent clogging of filters and valves;
Procedure:
1) Connect the machine to water connections (Fig. 9).
2) Connect the grey drain pipe to the drain connection
located at the back of the machine.
The correct positioning and sizing of the drain trap are
shown on the installation drawing (see the Installation
Plan).
Ö
WARNING! The washer disinfector drain pipe
must be maintained if backflow and other
drainage issues are to be avoided. Keep the
drain pipe free from kinks and knots, and
minimize the number of twists and turns in
the pipe. Be careful not to crush the pipe.
If the pipe has been crushed, kinked, or
otherwise damaged, consider replacing it to
prevent future drainage issues.