background image

11

Operative Instructions

understood the information in this manual, as well as 

by  unskilled personnel who  are  not  in good  mental 

and physical health.

• The machine must not be operated with the guards 

removed or partly damaged.

• Use  the  washer  disinfector  only  for  the  operations 

described in this manual.

• Do not damage or change the power cable or plug.

• Never  start  the  machine  if  the  cable  or  plug  are 

damaged.

• Do not pull the power cable to unplug it. Always act 

on the plug.

• Use  detergents  and  additives  specific  for 

manufacturer-approved 

washer 

disinfectors. 

Always  observe  the  manufacturer’s  instructions.  If, 

despite this, the product has negative effects on the 

instruments or machine, the responsibility will be of 

the manufacturer of the cleaning liquids.

• Only  introduce  instruments  that  can  be  treated 

with  automatic  cleaning  and  disinfection  process 

(see  manufacturer’s  instructions).  It  is  particularly 

important  to  follow  the  manufacturer’s  instructions 

when inserting new instruments that are used for the 

first time.

• If  additional  accessories  are  used  to  load  the 

instruments, especially hollow ones, the instructions 

contained  in  the  manufacturer’s  instruction  manual 

must be observed.

• Take care not to injure yourself by arranging sharp or 

pointed objects vertically. Try to arrange them in such 

a way as to prevent anyone from injuring themselves.

• Handle  the  chemical  products  with  care.  Chemicals 

for cleaning, neutralizing and rinsing contain irritants 

and caustic substances.

• The water in the chamber is not drinkable.

• Do not lean on the door and do not use it as a step.

• The  machine  during  its  work  cycle  could  reach  a 

temperature of 203°F (95°C); be very careful: there 

may be a risk of scalding.

• Do not change, for any reason, the characteristics of 

the  appliance,  its  installation  specifications  and  the 

parameters set.

• At the end of loading and unloading the instruments 

to be used, always close the door in order to avoid 

possible unpleasant smells coming from the drain.

• In case of fire, to extinguish the flames intervene with 

a powder fire extinguisher, DO NOT USE WATER.

• Do  not  wash  the  machine  with  direct  or  pressure 

water jets, or corrosive substances.

• Do  not  use  the  machine  to  wash  objects  and/or 

containers that, because of their shape or material, 

are not compatible with the indications given by the 

manufacturer. For objects to be washed, please follow 

the instructions explicitly indicated in this manual.

• In case of long outage periods of the machine, please 

cut the power supply off and close the water taps.

• Do  not  try  to  open  the  chamber  door  during 

operation:  the  appliance  is  equipped  with  a  special 

safety  lock  system  to  prevent  the  door  from  being 

opened.

 

d

HAZARD! Pay attention where indicated, for a 

potential danger of hot surfaces highlighted 

on the machine with this pictogram.

 

c

HAZARD!  Pay  the  utmost  attention  where 

indicated,  for  a  potential  electrical  hazard 

highlighted  on  the  machine  with  this 

pictogram.

3.1.4  Maintenance service personnel obligations

• Periodically check the integrity of the machine as a 

whole and the protection devices.

• Respect the laws in force in the country of use of the 

machine,  in  relation  to  the  use  and  disposal  of  the 

products used for cleaning and maintenance. Dispose 

of any special waste through appropriate companies 

authorised for this purpose, with issue of a receipt of 

the successful disposal.

• The assembly of parts of other brands or any changes 

(in  addition  to  voiding  the  warranty),  can  vary  the 

machine characteristics and, therefore, compromise 

its operational safety.

• If  the  protective  casings  are  removed,  make  sure 

that  they  are  correctly  restored  before  reusing  the 

machine.

• At the end of the maintenance and repair operations, 

before  restarting  the  machine,  make  sure  that  the 

work is completed, the safety devices reactivated and 

the guards reassembled.

• It is strictly forbidden to remove or tamper with the 

safety devices.

• The machine maintenance must only be performed 

with the power supply off, by qualified personnel and 

following the instructions in this manual.

4  Handling

Usually  the  packaged  and  palletised  machine  is 

transported  to  the  retailer/dealer  who,  by  means  of 

its personnel and suitable means, in compliance with 

current regulations, will itself make the delivery to the 

end user, ensuring transport and unloading operations 

depending on the type of transport vehicle. 

Summary of Contents for TIVA8-MD

Page 1: ...perate with the washer disinfector read these instructions for use INSTRUCTIONS FOR USE Original instructions Code Man205 0745009EN Rev 02 Edit 06 2022 Language English Firmware version 00B03 WASHER D...

Page 2: ...due to negligence and failure to comply with this manual and is also not liable for any damage caused by incorrect interpretations of the instructions contained Check for any updates to this manual o...

Page 3: ...6 2 5 Treatment of ophthalmic instruments 24 6 3 Programs 24 6 3 1 Suggested programs vs configurations 24 6 3 2 Program structure 24 6 4 Starting the machine 24 6 4 1 Before starting the program 24 6...

Page 4: ...r features The same personnel must store it and keep it intact to allow its consultation throughout the life span of the machine itself The personnel in charge must strictly and diligently follow the...

Page 5: ...environment with a maximum humidity of 80 for temperatures up to 88 F 31 C decreasing linearly to 50 at 104 F 40 C 2 2 Label specifications Thewasherdisinfectorisequippedwithanidentification label for...

Page 6: ...ance and assistance of the product contained in this use and maintenance manual have not been complied with If chemicals not identified by the manufacturer as compatible with the product are used With...

Page 7: ...he power supply voltage on the line neutral 1200V and a duration up to 5s of long duration which may occur between the line conductor and the system earthing which may have a voltage equal to the powe...

Page 8: ...oise during wash phases 56 dB A Max dBA noise during drying phase 61 dB A Type of protection according to IEC 60529 IP20 IP International Protection Against ingress of solid foreign objects 12 5 mm di...

Page 9: ...level basic version 3b Washing rack with spray arm positioned at the upper level basic version 3c 1 1 DIN basket manufactured in stainless steel suitable for washing all kinds of instruments 3d Suppor...

Page 10: ...zed people from using the machine The user undertakes to adequately inform and train its personnel on the application and observance of the safety regulations Failure to comply with these rules may je...

Page 11: ...ER Do not wash the machine with direct or pressure water jets or corrosive substances Do not use the machine to wash objects and or containers that because of their shape or material are not compatibl...

Page 12: ...operation takes place and check the integrity and suitability of the lifting and transportation equipment available Also make sure that the area of operation is clear and that there is sufficient esc...

Page 13: ...ills 4 1 Storage If the packaged machine is temporarily stored make sure that it is not subjected to blows and tampering However it must be placed in a closed dry dust free environment and protected f...

Page 14: ...supply in compliance with the current laws and regulations a WARNING In the single phase version the main switch of the power supply line must be a multi pole circuit breaker with adequate residual cu...

Page 15: ...e disconnection procedure of the electrical connection by placing the dedicated safety device in OFF state The upstream electrical power line must be dimensioned and protected in accordance with curre...

Page 16: ...IG 500 kPa 72 PSIG 500 kPa Connect the cold hot and deionized if available water hoses exiting the machine with their respective network connections as shown on the installation drawing It will be the...

Page 17: ...ing the appropriate funnel a WARNING During this operation check that the plastic cap is closed 4 Insert the rack and start a normal washing cycle The machine regenerates automatically a WARNING The w...

Page 18: ...ed manually regardless of the warning message on the display 5 4 Connection of the drain trap A B 1 1 3 2 2 Fig 10 Water and drain system Fig 10 A Floor drain connection B Wall drain connection 1 Hose...

Page 19: ...ter 4 Salt tank cap Insert the supplied filters in the specific positions Constantly check the cleanliness of the filters especially the chamber bottom filter 3 Fig 11 Use this filter to have a high f...

Page 20: ...ren and outsiders Possibly locked away Use on only the manufacturer s approved products Do not use liquids for household dishwashers 5 6 1 Refill procedure 1 Set up a new tank with the chemical 2 Remo...

Page 21: ...be cleaned a WARNING Check that there are no water leaks in the pipe fittings 2 Check that the suction tubes located in the lower part of the machine within the corresponding tanks containing the liqu...

Page 22: ...the door cannot be opened normally d HAZARD If the emergency release is used while a program is running very hot water and chemicals may leak Therefore there is a danger of burns scalding and irritat...

Page 23: ...e baskets Before the automatic treatment check that the lumen and hollow body instruments can be treated internally and remove any obstructions Carefully place the load in the supports Objects must no...

Page 24: ...ne leaves the factory with presetting washing programs already entered in the settings menu For programs details see the Washing Programs booklet 6 3 1 Suggested programs vs configurations For a good...

Page 25: ...too high water temperatures Wash It is the actual cleaning cycle The cleaning chamber heats up until it reaches the specific temperature for the selected program a temperature that remains stable for...

Page 26: ...burns a WARNING Failure to follow these recommendations can cause burns 7 Control Panel 1 2 3 4 8 7 6 5 Fig 15 The control panel consists of 8 keys and a 3 5 inch LCD display All keys except key 3 ar...

Page 27: ...the number of the alarm and a brief description 3 Water conductivity value in microsiemens if conductivity probe function is enabled 4 Several animated images showing the current status of the machine...

Page 28: ...and close the door WARNING The door must be tightly closed until you hear the classic closing click otherwise the program will not start Fig 18 1 2 With the machine started and door closed the screen...

Page 29: ...t in the chamber Fig 18 6 Phase 2 6 Self loading of cold water During the self loading phase the chamber will be filled Fig 18 7 7 Prewash phase begins Fig 18 8 Phase 3 8 At the end of the prewash pha...

Page 30: ...t Fig 18 12 Phase 5 12 At the end of the wash phase the water is automatically discharged Fig 18 13 Phase 6 13 Self loading of hot water During the self loading phase the chamber will be filled Fig 18...

Page 31: ...of the water At the end of the phase if the conductivity will be below a presetted value the machine will procede to the next phase If the check is negative the machine will repeat phases 7 and 8 aga...

Page 32: ...control panel and a red led light in the chamber At first the image concerning the alarm is displayed for 5 seconds then the alarm description is displayed for 10 seconds The image and text alternate...

Page 33: ...consists of two screens the item when selected lights up Fig 20 3 Fig 20 4 7 2 7 Settings See Annexes 12 1 and 12 2 7 2 8 Language setup To change the display language with the door open press PRG fo...

Page 34: ...this software in fact it s possible to create and print digital versions of the reports connect the device to the PC to customize the user names and passwords via USB port cable not included use USB...

Page 35: ...ater The value increases by 1 C at a time ERR Alarms 0000 means that the machine is not in an alarm state When an alarm is triggered its code will be transmitted until the alarm is reset CYC Programs...

Page 36: ...in this manual a WARNING The washer disinfector maintenance operations must be carried out with the machine completely switched off a WARNING If the machine is not used for more than 24 hours it is ne...

Page 37: ...brushes bleach bleach cause the oxidation of the stainless steel surface and consequent change of color or any cleaning agent containing bleach or anything abrasive to cleaning the device Doing so wil...

Page 38: ...WARNING It is mandatory to replace the prefilter and the HEPA H14 filter as indicated in the Maintenance Logbook In case of intensive use of the machine it is recommended to increase the frequency of...

Page 39: ...are not disposed of correctly or are disposed of as household waste these components can damage human health and the environment Never dispose of the old washer disinfector among conventional waste U...

Page 40: ...isables hot water If hot water is enabled purified water is automatically deactivated and vice versa OFF 8 User PIN enable at the start of cycle 0 OFF 1 ON Available only with User PIN Management opti...

Page 41: ...Set password expiration time 6 22 User PIN enable at the end of cycle 0 OFF 1 ON Available only with User PIN Management option Enables or disables user PIN at the end of cycle OFF 23 Upper Spray Arm...

Page 42: ...N Enables or disables the product 2 flowmeter ON 11 Product 3 enable 0 OFF 1 ON Enables or disables the product 3 pump ON 12 Product 3 flowmeter 0 OFF 1 ON Enables or disables the product 3 flowmeter...

Page 43: ...program 6 Program cycles 40 Number of cycles performed with program 40 USER PIN 1 Pin User 1 To set the user password press the RESET key and enter the 6 desired numbers 2 Pin User 2 3 4 Pin User 20...

Page 44: ...y language Resuming mode Program bu on 1 Program bu on 2 Program bu on 3 Printer language Drying enable Lower Spray Arm White LED always ON Fahrenheit temperature PIN expira on month User PIN enable a...

Page 45: ...password change D O P S 3 Select User User Blocked Press Start to con rm P D P D P D P D P D P D Y N N Y N Enter old password Enter new password Con rm new password The password has been changed corr...

Page 46: ...r user S P P 5 Select User Pin menu Insert Super User password max 3 a empts Digit password is valid is expired or rst access See password change procedure for Super User P D P D P D N N Y Y Select Us...

Page 47: ...word expired Press Start to con rm N N Enter old password Enter new password Con rm new password The password has been changed correctly Is valid Is it equal to old password Y Y N Y S P P 5 Insert Sup...

Page 48: ...WATER The hot water tap could be closed or partially open Check it Press RESET If it persists call the service personnel E22 NO DEMI WATER The demineralized water tap could be closed or partially open...

Page 49: ...ight be broken or disconnected Press RESET If it persists call the service personnel E53 CHAMBER T DIFF A temperature difference between the probes in the chamber has been detected too high Press RESE...

Page 50: ...D RESERVE PRODUCT 1 Product 1 is in reserve or may be empty Add the product 1 LIQUID RESERVE PRODUCT 2 Product 2 is in reserve or may be empty Add the product 2 LIQUID RESERVE PRODUCT 3 Product 3 is i...

Page 51: ...s not reached The thermostat sensor in the washing chamber is covered with deposits Clean the thermostat sensor 5 The machine does not dry The air filter in the drying system is dirty or clogged Repla...

Page 52: ...PT1000 2 temperature chamber control probe C Yes 0 3 1 0 31 0 17 18 19 20 TPT3 PT1000 3 temperature air probe C Yes 0 1 0 4 104 21 22 23 24 ERR Alarm number n Yes 0 0 7 0 70 25 26 27 28 29 30 CYC Coun...

Page 53: ...0 0 6 6 71 72 73 74 CONP Conductivity probe status n Yes 0 0 0 2 2 75 76 77 78 79 A0 A0 Value n Yes 0 6 1 8 0 6180 80 81 82 83 SPRYR B Spray arm at the bottom of the chamber rotation monitoring rpm Y...

Page 54: ...0 Program 40 PHA TYPE PHASE IN EXECUTION 0001 Drain 0010 Trolley insertion 0002 Prewash 0011 Door closure 0003 Wash 0012 Program selection 0004 Rinse 0013 Operator scan 0005 Disinfection 0014 Trolley...

Page 55: ...150 150 ml of product delivered VCW VHW VDW QUANTITY OF WATER DELIVERED 0000 0 liters of water delivered liters of water delivered 0001 1 liter of water delivered 0012 12 liters of water delivered CON...

Page 56: ...56 Operative Instructions Notes...

Page 57: ...57 Operative Instructions Notes...

Page 58: ...58 Operative Instructions Notes...

Page 59: ...59 Operative Instructions...

Page 60: ...Tuttnauer USA Co Ltd 25 Power Drive Hauppauge N Y 11788 USA Tel 1 631 737 48 50 1 800 624 58 36 Fax 1 631 737 07 20 Web www tuttnauerusa com...