42
Operative Instructions
ID Parameter Name
Min
Max
Description
Factory
Default
4
Hot water filling
time
0
1000
Sets the timer for the pump in the
absence of the flowmeter count
52
seconds
5
Deionized water
flowmeter
0=OFF
1=ON
Enables or disables the deionized water
flowmeter
ON
6
Deionized water
filling time
0
1000
Sets the timer for the pump in the
absence of the flowmeter count
52
seconds
7
Product 1 enable
0=OFF
1=ON
Enables or disables the product 1 pump ON
8
Product 1
flowmeter
0=OFF
1=ON
Enables or disables the product 1
flowmeter
ON
9
Product 2 enable
0=OFF
1=ON
Enables or disables the product 2 pump OFF
10 Product 2
flowmeter
0=OFF
1=ON
Enables or disables the product 2
flowmeter
ON
11 Product 3 enable
0=OFF
1=ON
Enables or disables the product 3 pump ON
12 Product 3
flowmeter
0=OFF
1=ON
Enables or disables the product 3
flowmeter
ON
13 Product 4 enable
0=OFF
1=ON
Enables or disables the product 4 pump OFF
14 Product 4
flowmeter
0=OFF
1=ON
Enables or disables the product 4
flowmeter
ON
REGENERATION
1
Water hardness
(°fH)
0
60
French degrees of water.
1°fH = 10 PPM CaCO
3
25°fH
CONDUCTIVITY PROBE
1
Probe activation
0=OFF
1=ON
Enables or disables the conductivity
probe
OFF
CLOCK
1
Date and Time
DW DD/MM/YY
HH:MM:SS
Sets current date and time
CYCLES COUNTER
1
Total Cycles
Number of machine life cycles
2
Program cycles 1
Number of cycles performed with
program 1
3
Program cycles 2
Number of cycles performed with
program 2
4
Program cycles 3
Number of cycles performed with
program 3
Summary of Contents for TIVA8-MD