50
Operative Instructions
ALARM
DESCRIPTION
ACTION
E78 PRODUCT 4 EMPTY
Chemical liquid tank 4 is empty.
Add liquid.
Press RESET
If it persists, call the service personnel.
E80 SPRAY ARM LOCKED
The spray arm marked in red on
display is locked. Remove it and
clean.
Press RESET
If it persists, call the service personnel.
E89
NO DISINFECTION
The minimum value of A0 was
not reached and the load is still
contaminated.
Press RESET
If it persists, call the service personnel.
12.7 Warnings table
WARNING TYPE
DESCRIPTION
ACTION
LIQUID RESERVE
PRODUCT 1
Product 1 is in reserve or may be empty.
Add the product 1
LIQUID RESERVE
PRODUCT 2
Product 2 is in reserve or may be empty.
Add the product 2
LIQUID RESERVE
PRODUCT 3
Product 3 is in reserve or may be empty.
Add the product 3
LIQUID RESERVE
PRODUCT 4
Product 4 is in reserve or may be empty.
Add the product 4
REQUEST
MAINTENANCE
Contact customer service or the dedicated
technical assistance for periodic maintenance.
Press RESET
REQUEST
VALIDATION
Contact customer service or the dedicated
technical assistance for periodic validation.
Press RESET
REFILL
SALT TANK
1- Take out the basket
2- Unscrew the cap of the salt tank
3- Fill the salt tank
4- Screw the cap of the salt tank
Press RESET
For 5 seconds