Page 9
cleaning, prevention of clogging and preservation from corrosion,
what is permitted and what is prohibited for sterilization and choosing
a sterilization program.
4.
Liquids may be sterilized only with the “liquids” program. The
container must be covered but not sealed. Sealed bottles may only be
sterilized using a special program. The bottle must be either Pyrex or
a Borosilicate glass bottle.
5.
When sterilizing plastic materials, make sure that the item can
withstand sterilization temperature. Plastic that melts in the chamber
is liable to cause a great deal of damage.
6.
Individual glass bottles may be placed within an appropriate container
that will be placed on a tray. Never place glass bottles on the floor of
the autoclave. Never fill more than 2/3 of the bottle volume.
7.
On closing the autoclave's door, make sure it is properly locked
before activating.
8.
Before withdrawing trays, wear heat resistant gloves.
9.
Before opening the door, verify that there is no pressure in the
chamber.
10.
Open the door slowly to allow steam to escape and wait 5 minutes
before you remove the load. When sterilizing liquids, wait 10 minutes.
11.
Once a month, ensure that the safety valves are functioning, and
once annually a certified tester must conduct pressure chamber
safety tests.
12.
Once annually, or more frequently, effective tests must be performed,
i.e., calibration and validation.
13.
Examine the condition of assemblies on a regular basis. Make sure
there are no leaks, breaks, blockages, whistles or strange noises.
14.
It is required to conduct maintenance operations as instructed.
15.
Immediately notify the person in charge of any deviation or risk for the
proper function of the device.