Page 8
1.6
Accessories
•
Baskets
– Different size baskets are available for these units. The
baskets are made of stainless steel wire and have a handle. The
basket allows the operator to load a large quantity of materials into the
chamber.
The baskets are made of stainless steel wire and have a handle. The
basket allows the operator to load a large quantity of materials into the
chamber.
•
Stainless steel containers - Different size containers are available for
these units .
The containers are designed for sterilizing waste material. The
containers have vent holes along the upper rim.
2
Safety Instructions
The autoclave has unique characteristics. Please read and understand
the operation instructions before first operation of the autoclave. The
following issues may require instructions guidance provided by the
manufacturer: how to operate the autoclave, the door safety
mechanism, the dangers involved in circumventing safety means, how
to ensure that the door is closed, and how to select a correct
sterilization program.
Make sure that you know where the main power switch is, where the
water cut-off valve is and where the compressed air disconnection
valve is located.
Autoclave maintenance is crucial for the correct and efficient function of
the device. We enclose a log booklet that includes maintenance
recommendations, with every device.
The weekly spore test is part of the preventive maintenance plan, along
with the annual validation of the sterilization processes that ensures
appropriate temperature dispersion within the chamber.
Never use the autoclave to sterilize corrosive products, such as: acids,
bases and phenols, volatile compounds or solutions such ethanol,
methanol or chloroform nor radioactive substances.
1.
Never start using a new autoclave or a new steam generator, before
the safety, licensing and authorization department has approved it for
use.
2.
All autoclave users must receive training in proper usage from an
experienced employee. Every new employee must undergo a training
period under an experienced employee.
3.
A written procedure must be established for autoclave operation,
including: daily safety tests, seal inspection and door hinge
inspection, smooth action of the closing mechanism, chamber