Page 89
Problem/
Error
Message
Message /
Symbol Description
Corrective Action
"RTC Error -
Please Set
Current
Date and
Time"
This message is displayed in order
to set the date and the time.
Set Current Date And Time.
If the problem persists, call
the technician.
"Time Error"
This message is displayed if the real
time clock is faulty.
Call the technician.
"Door is
open
(During the
cycle)"
This message is displayed when the
door is open: During the cycle.
Close the door to perform a
new cycle.
"Canceled
By User"
This message is displayed after the
START/STOP key is pressed and
cycle aborted.
Wait until "cycle failed
–
canceled by user" or "cycle
end
– canceled by user" is
displayed.
Perform a new cycle.
"Cycl
e Failed"
This message and symbol are
displayed if an error occurs before
sterilization cycle is completed.
Perform a new cycle.
"Air Error"
This message is displayed at the
end of the cycle If the autoclave
does not reach the atmospheric
pressure after 10 minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a new
cycle.
Compressed
air supply
error
PV(pre-
vacuum
models only)
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the
compressed air supply.
"Periodical
check time
exceeded -
Please call
for service"
The periodical maintenance time has
passed.
Call for service.
"Mineral
free water
reservoir
empty"
This message is displayed if the
water level electrode does not sense
water.
Fill the mineral free water
reservoir.
"Cycle
counter
exceeded -
Please call
for service"
Number of cycles, since last
periodical maintenance, exceeded
the preset number as defined by
"cycle counter" parameter.
Call for service.
"Power
Down"
This message is displayed if power
down has occurred during the cycle.
(this message will print out in the
Turn on the autoclave and
wait until the autoclave is
ready (reaches the safe