If the autoclave is equipped with automatic mineral-free water supply,
this instruction applies to the clean-water reservoir only. In this case
close the mineral-free water supply valve before draining the reservoir.
1.
To drain the reservoir, use item (5) with the plastic hose (6) attached to
it (supplied with the autoclave).
2.
Insert part (5) into valve (3) and press it until you hear a “click”. The
drain valve is open.
3.
When the water reservoir is empty, press part (4). Item (5) will pop out
approx. 3mm and the drain valve will be closed. Remove item (5) with
the plastic tube.
4.
If the drained reservoir is the clean-water reservoir, fill reservoir with
distilled water until, reaches full level. (approximately 6.5 liters (1.72
gallons)).
The autoclave is now ready for use.