background image

 

Page 40 

 

 

Slow Exhaust:

 Steam is exhausted from the chamber at a slow rate, 

until it reaches the pressure of 30 kPa above the ambient pressure. 
Then  steam  is  exhausted  out  of  the  chamber  at  a  fast  rate  until 
pressure decreases to ambient pressure.

   

Note

:  If  the  cycle  fails,  fast  exhaust  is  performed  through  the  bio -hazard 

filter. Some water may remain in the chamber after a failed cycle. To 
remove such water, press Start and repeat the cycle. If the problem 
persists, call for service. 

 

Drying:

 N/A 

 

Ending:

 The system checks that End temperature and End pressure 

have been reached, and then releases the door.  

 

5.12

 

Program 12: Vacuum Test (PV, pre-vacuum models 
only) 

 

 

 

 

Vacuum Test is a test program with the following parameters: 

 

 

Nominal parameters default settings 

 

Vacuum pressure: 17.0 kPa 

 

Vacuum time 1: 5 minutes 

 

Vacuum time 2: 10 minutes 

 

Operations sequence: 

 

Vacuum  is  produced  in  the  chamber  down  to  P1=17  kPa.  At  this 
stage all the valves close. The autoclave remains in this stage for 5 
minutes.  This  period  enables  the  condition  in  the  chamber  to  reach 
equilibrium.  

 

After  the  5  minutes  have  elapsed,  the  printer  records  the  pressure 
that  is  referred  to  as  P2.  At  this  point  the  test  begins  and  lasts  10 
minutes.  

 

At the end of the test, the printer records the results. The pressure at 
the end of the test is referred to as P3. The rate of change of P3-P2 
shall not exceed 0.13 kPa/min.).  

 

Summary of Contents for LABSCI 11L CPVG

Page 1: ...ELG D standard autoclave C optional system for fast cooling PV optional vacuum pump BH optional bio hazard filtration F optional fan for super fast cooling Cat No MAN205 0480002EN Rev H Heidolph Nort...

Page 2: ...ll Dimensions 14 3 9 Electrical Data 14 3 10Utilities 15 3 11Environment Emission Information 15 3 12Construction 15 3 13Safety Features 16 3 14Stickers Description 16 3 15Front View 18 3 16Rear View...

Page 3: ...and Dick Test PV pre vacuum models only 41 5 14Program 14 Warm Up 42 5 15Program 15 Isothermal 43 5 16Program 16 Air Mixture 44 5 17Program 17 Glass Test 45 5 18Program 18 Durham 46 6 Screens 48 6 1...

Page 4: ...11 3Quick options screen 69 11 4Logging in and entering the Main menu 73 11 5Changing parameters 74 11 6System Parameters 75 11 7Maintenance 76 11 8Cycle parameters 77 12 Service and Maintenance Inst...

Page 5: ...ng in effective sterilization Better temperature uniformity Better drying of materials with closed doors due to the vacuum achieved in the chamber at the end of the sterilization cycle In BH bio hazar...

Page 6: ...is provided with a pressure gauge that is used as guide only Should there be a power failure during the operation of the autoclave the pressure gauge indicates to the operator that there is pressure...

Page 7: ...certify that this instrument is guaranteed to be free from defects in material and workmanship for one year against faulty components and assembly with the exception of glassware lamps and heaters Th...

Page 8: ...instrument and the solution is not covered in this manual contact our representative or us first Do not attempt to service this instrument yourself Describe the difficulty as clearly as possible so w...

Page 9: ...program Make sure that you know where the main power switch is where the water cut off valve is and where the compressed air disconnection valve is located Autoclave maintenance is crucial for the co...

Page 10: ...r fill more than 2 3 of the bottle volume 7 On closing the autoclave s door make sure it is properly locked before activating 8 Before withdrawing trays wear heat resistant gloves 9 Before opening the...

Page 11: ...ver 2000 meters 6561 feet ambient pressure shall not be lower than 80 kPa 11 6 psi The autoclave shall not be used in a manner not specified in this manual Do not use the autoclave in the presence of...

Page 12: ...61010 2 040 Safety IEC IEC 61326 1 EMC ISO 17665 1 2006 Validation and Routine Control 3 4 Electrical Data Property Value Total Power 9000W Voltage 230VAC 1ph Amperage 10A Protection against electrica...

Page 13: ...0 1 mg l Conductivity at 25 C 3 to 15 s cm 3 s cm pH value degree of acidity 5 to 7 5 5 to 7 Hardness ions of alkaline earth 0 02 mmol l 0 02 mmol l Appearance Colorless clean without sediments Compl...

Page 14: ...m Length 760mm Maximum dimensions door open Width 1090mm Distance between supporting legs Width 315mm front legs 400mm rear legs Length 532mm Weight per support area max load According to overall weig...

Page 15: ...tion against electrical shock Class I IEC 60601 1 Mains supply fluctuation 10 Degree of protection by enclosure IP31 Note In order to avoid any injury by electrical hazard it is recommended that a gro...

Page 16: ...ipment within easy reach of the operator and marked as the disconnecting device for the equipment The electrical net must be protected with a current leakage safety relay The electrical network must c...

Page 17: ...peration to be started only when the door is closed Water level safety device Excess temperature protection 3 14 Stickers Description Symbol Meaning Part Number Location Caution Hot steam LAB048 0058...

Page 18: ...Page 17 ON OFF On Off LAB048 0018 Near the Power switch...

Page 19: ...ption 1 Main switch circuit breaker 10 Door switch 2 Operating keyboard 11 Door closing device 3 Printer cover 12 Water reservoir funnel 4 Rear Leg 13 Water level gauge 5 Printer option 14 Autoclave c...

Page 20: ...cover 3 16 Rear View No Description No Description 1 Water reservoir cover 7 Tap water inlet 2 Compressed air inlet 8 Overflow drain 3 Compressed air pressure gauge 9 Chamber water strainer 4 Air pre...

Page 21: ...Page 20 4 Control Panel 4 1 Control Panel Drawing No Description 1 Display 2 Keypad Up Button 3 Keypad Start Stop Button 4 Keypad Down Button 5 Printer 3 5 1 2 4...

Page 22: ...d below UP key This key has the following functions In the menu directories o This key enables the operator to browse through the cycles In the directories available o When the cursor is blinking on a...

Page 23: ...the ERROR message displayed on the screen and opens the electric door lock In the menu directories o When the cursor is blinking on a number the START STOP key enables moving to the next position o W...

Page 24: ...y Clock mode is performed Enter the Admin menu as described in this manual to change the time or to cancel this option Cycle by clock This message is displayed if the user presses START STOP key while...

Page 25: ...reached the timed sterilization cycle begins 4 5 4 Cooling Optional The autoclave is designed to operate two liquid cooling cycles as follows Sealed bottles cooling with compressed air On completion...

Page 26: ...ssure at the end of the test is referred to as P3 The rate of change of P3 P2 shall not exceed 0 13 kPa min If P3 P2 exceed 0 13 kPa min the printer will print FAIL If P3 P2 is within the required val...

Page 27: ...Time minutes C models only PV models only BH models only Icon No Name Glass 134 C 273 F 3 1 Plastic 121 C 250 F 15 1 Liquid A 121 C 250 F 15 Liquid B Waste 121 C 250 F 30 Liquid A Cooling 121 C 250 F...

Page 28: ...F 45 Vacuum test PV only 5 1 0 Bowie and Dick 134 C 273 F 3 5 1 Warm Up 80 C 176 F 20 Isotherm al 80 C 176 F 20 Air Mixture 121 C 250 F 15 Glass Test 121 C 250 F 20 Durham 121 C 250 F 15 These progra...

Page 29: ...Page 28 The stages names are as follows Start Pulse Low PV models only Pulse High PV models only Heating Sterilization Cooling C models only Exhaust Drying PV model only Ending...

Page 30: ...ly Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reached Sterilization Sterilization tem...

Page 31: ...m models only Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reached Sterilization Steril...

Page 32: ...s only Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reached Sterilization Sterilization...

Page 33: ...e vacuum models only Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reached Sterilization...

Page 34: ...pre vacuum models only Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reached Sterilizati...

Page 35: ...sequence PV pre vacuum models only Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reache...

Page 36: ...perations sequence Pulse low Pulse high four vacuum pulses remove air by drawing the chamber down to 25kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reach...

Page 37: ...s Drying time 15 minutes Operations sequence Pulse low Pulse high four vacuum pulses remove air by drawing the chamber down to 25 kpa Heating Steam enters the chamber and heats it up until the sterili...

Page 38: ...high at one pulse it will build vacuum down to 25 kPa All exhaust from the chamber is performed through the bio hazard filter Heating Steam enters the chamber and heats it up until the sterilization t...

Page 39: ...ons sequence PV pre vacuum models only Pulse low Pulse high at one pulse it will build vacuum down to 25 kPa All exhaust from the chamber is performed through the bio hazard filter Heating Steam enter...

Page 40: ...rature sensors must be inside the bottles For proper sterilization fill the bottles with approximately the same amount of liquid Nominal parameters default settings Sterilization temperature 121 C 250...

Page 41: ...eases the door 5 12 Program 12 Vacuum Test PV pre vacuum models only Vacuum Test is a test program with the following parameters Nominal parameters default settings Vacuum pressure 17 0 kPa Vacuum tim...

Page 42: ...nce Pulse low Pulse high Four vacuum pulses will build vacuum down to 25 kPa Heating Steam enters the chamber and heats it up until the sterilization temperature is reached Sterilization Temperature a...

Page 43: ...0 minutes Operations sequence Pulse low Pulse high N A Heating Steam enters the chamber and heats it up until the warm up temperature is reached Warm up Warm up temperature is maintained constant for...

Page 44: ...ion program Put one PT100 inside the bottle leave the second one hanging in the chamber outside the bottle see below Nominal parameters default settings Heating temperature default 80 C 176 F Heating...

Page 45: ...inutes Sterilization is controlled in a way that the sterilization pressure remains approx 30 above the theoretical pressure corresponding to the same temperature according to the steam table These in...

Page 46: ...checks that End temperature and End pressure have been reached and then releases the door 5 17 Program 17 Glass Test This cycle is intended to check the durability of the bottles for liquids when the...

Page 47: ...ge of the cooling take approximately 21 minutes This is to check the durability of the bottles Fast Exhaust The steam is exhausted out of the chamber at a fast rate until pressure decreases to ambient...

Page 48: ...ization time Slow Exhaust Steam is exhausted from the chamber at a slow rate until chamber temperature reaches 105 C Cooling Forced cooling to the required end temperature chamber pressure remains at...

Page 49: ...6 1 Screens following a completely successful cycle Cycle Ended 1 System Ready 2 Pulse Low PV model only 3 Pulse High 1 1 PV model only 5 Heating 6 Sterilization 7 Exhaust 8 Drying PV model only 9 End...

Page 50: ...erilization stage The cycle ended successfully the reason for aborted cycle is displayed 6 2 2 Pressure Time Error Failure occurrence after complete sterilization stage The cycle ended successfully th...

Page 51: ...age When Cycle Failed appears on the screen the user shall press START STOP key in order to delete the Cycle Failed message An example for all displayed warnings according to Cycle Failed Cycle Failed...

Page 52: ...e date and ending with Cycle Ended For an aborted cycle Cycle Failed and the Error message are printed refer to Displayed Error Messages Symbols For an example of a typical printout see next page Note...

Page 53: ...ture and pressure during heating A 00 08 48 089 6 077 1 The time temperature and pressure during air removal A 00 05 48 091 3 049 4 The time temperature and pressure during air removal W 00 03 22 092...

Page 54: ...moval stage CLK 2 Software clock H Heating stage E Exhaust stage K Keep Heat Optional D Drying Stage ELPV Model only C Cooling stage for ELVC only 7 2 Printer Handling 7 2 1 Maintenance Wipe off the s...

Page 55: ...s shown see Fig 3 2 4 Take out the paper and re close the cover as shown see Fig 3 3 the printer cover is locked 5 Tear off the exceeding paper using the jagged edge see Fig 3 4 Fig 3 6 Close the prin...

Page 56: ...liquids If any liquids get into the printer it could lead to fire electric shock or other serious accidents Never touch the thermal head immediately after printing because it becomes very hot Make su...

Page 57: ...o allow ventilation and to facilitate the device disconnection If placed in a cabinet verify that the rear of the cabinet is open to allow ventilation Insufficient space for ventilation may result in...

Page 58: ...h instruments with an ultrasonic cleaner using detergent and mineral free water 4 Launder textile wraps prior to reuse 5 After cleaning rinse instruments for 30 seconds Follow manufacturer s instructi...

Page 59: ...uate separation of packaged instruments 9 Load trays in such a way as to allow steam to move freely among all items 10 Once a week use a biological spore test indicator in any load to make sure steril...

Page 60: ...packs of folded operating room drapes with layers vertical allowing air to be removed from the packs rapidly 6 Do not place packs of hollowware and trays of instruments above textile packs or soft goo...

Page 61: ...ressure build up 3 The PT100 temperature sensors se are used to control the program temperature and ensure the safety of the operating cycle Note For isothermal cycles insert one PT100 temperature sen...

Page 62: ...oclave 10 2 Opening the door 1 Place your thumb on the plastic door cover 1 and the other fingers in the handle 3 2 Pull the handle 2 until the locking of the door is released 3 Open the door When the...

Page 63: ...5 2 10 4 1 Closing the door 1 Hold the opening handle in open position while pushing the door until it comes to closed position then release the handle 2 The open door symbol is replaced by the messa...

Page 64: ...oor is not completely closed the process will not start and the DOOR indicator will flash twice then turn off and the buzzer will sound four times Caution Do not touch the strainer s cover mounted on...

Page 65: ...are displayed on the screen Refer to Displayed Error Messages Symbols Pressing the START STOP key cancels the displayed message and enables opening the door If the door is not opened the vacuum pump...

Page 66: ...or detailed loading instructions see sec 9 Preparation before sterilization 10 9 3 Unloading On completion of the cycle take out the load immediately from the sterilizer Do not remove the load from th...

Page 67: ...u will learn how to browse through the folders When you read the Directories and subdirectories chapter with links to specific menus you will need to know how to browse through the folders using the a...

Page 68: ...owsed through the menus and reached the parameter changing screen as explained above Now you can change the required parameter as desired To do so 1 Enter the required value as follows Press the Up an...

Page 69: ...e note the maximum and minimum values for this parameter shown on the screen Your value must be within these boundaries Below is the example of changing the Dry time parameter on the screen used in th...

Page 70: ...P and DOWN keys simultaneously 11 3 Quick options screen When the autoclave is on and no cycle is running press Up and Down keys simultaneously to enter the Quick options screen Most of the options re...

Page 71: ...some Quick Options 11 3 1 Version Information This directory allows viewing information of the current factory default and previous software versions 1 Enter the Version information screen 11 3 2 Sta...

Page 72: ...ing the time to start the cycle 1 Move the cursor to the Time field 2 Set the required time Enabling the Start Cycle by Clock 1 Set the starting time 2 Move the cursor to Enabled Press Up or Down key...

Page 73: ...art Cycle by Clock 11 3 3 Set date and time This subdirectory enables the operator to set date and time On the Set date and time screen the time is displayed in the upper row in the form HH MM SS The...

Page 74: ...11 2 explains how to change a parameter When the autoclave is on and no cycle is running press the up and down keys simultaneously to enter the Quick Options screen see 11 3 On this screen you can ei...

Page 75: ...ptions available to user on the Main Menu 11 5 Changing parameters Bacsoft control panel provides an interface that consists of control screens available through an easy scrollable menu tree To learn...

Page 76: ...llowing folder Main menu System parameters You will see the following screen 11 6 1 Screen Saver In this menu you can define the screensaver delay time i e how long the keyboard will be untouched befo...

Page 77: ...for autoclave s maintenance menu 11 7 1 Reset atmospheric pressure In this menu you can reset the atmospheric pressure value To do so 1 Browse to the following folder Maintenance Reset atmospheric pr...

Page 78: ...done See the printout shown in the Printer handling chapter of the user manual 11 8 Cycle parameters The Cycle parameters menu includes parameters of a specific sterilization program cycle Browse to t...

Page 79: ...ameter as desired Drying stage is divided into two stages For each stage you can set the total time on time and off time So during the Dry First Stage the heating elements are on for the time set in D...

Page 80: ...y by the operator Clean door gasket with a soft cloth The gasket should be clean and smooth A mild soapy solution may be used 12 1 2 Weekly by the operator 1 Check the interior of the autoclave If the...

Page 81: ...ions Calibrate the temperature and pressure Perform validation of the autoclave Checking the precise operation of the earth leakage relay Checking that the autoclave is leveled Checking the safety ele...

Page 82: ...into valve 3 and press it until you hear a click The drain valve is open 3 When the water reservoir is empty press part 4 Item 5 will pop out approx 3mm and the drain valve will be closed Remove item...

Page 83: ...clean out wire inserted in it wire is permanently installed and will not come out It is required that the air jet be cleaned once per week or more often if necessary to remove any accumulated dirt and...

Page 84: ...steam pressure to escape through it every month To avoid injuries begin this check while the autoclave is cold 1 For the autoclave safety valve operate the sterilization cycle for the generator safety...

Page 85: ...Page 84 1 2...

Page 86: ...n the exhaust line during and shortly after operation Touching the hot strainer s cap may cause severe injuries 2 If maintenance operation is performed while strainer cap is hot use heat resistant glo...

Page 87: ...evel in the chamber is properly reported to the microprocessor all during the cycle The water sensor is located in the rear of the chamber It is easily cleaned using a damp cloth or sponge you may use...

Page 88: ...d Message is displayed and FAIL indicator lights if in the air removal stage a vacuum level of 25kPa is not reached during 20 minutes after the cycle is started Perform a new cycle Call the technician...

Page 89: ...Heat Time Error Keep This message is displayed if the system cannot reach the required temperature in the chamber during the optional Keep Heat stage within the preset time Verify that the autoclave i...

Page 90: ...ror This message is displayed at the end of the cycle If the autoclave does not reach the atmospheric pressure after 10 minutes Wait until the autoclave reaches the atmospheric pressure and perform a...

Page 91: ...the autoclave turns on condition and perform a new cycle No Water This message is displayed if the electrode in the chamber did not sense water within the preset time 1 check and fix the mineral free...

Page 92: ...NTAINERS Stainless steel wire baskets Stainless steel container for waste products L x D x H mm Capacity L x D x H mm Capacity 400 x 180 x 160 1 400 x 190 x 160 1 Pouch Rack ACS215 0008 Container for...

Page 93: ...Filter Air 0 2 Micron Model 50mm D THE002 0066 Thermal paper for CUSTOM PLUSII printer roll 57mm d 50mm 15 Accessories Part number Description GAS084 0007 Drain P V C Tube 8x12 THE002 0052 Printer PLU...

Reviews: