Message
Description
cannot reach the required temperature, in
the chamber, within the preset time.
"Low Pressure"
This message is displayed if Chamber
Pressure drops below the sterilization
pressure for 2 seconds during the
sterilization stage.
"High Pressure"
This message is displayed if Chamber
Pressure raises 0.29 bar above
sterilization pressure for 2 seconds
during the sterilization stage.
"High Pressure (Ending)"
This message is displayed if the system
cannot reach atmospheric pressure ±
0.05 bar during the ending stage.
"High Pressure (Exhaust)"
This message is displayed if the system
cannot reach preset pressure within 10
minutes from the beginning of the
exhaust stage.
"Pressure Time Error"
This message is displayed if the system
cannot reach the required pressure
conditions in the chamber, after preset
time, during the air removal stage.
"Time Error"
This message is displayed if the real time
clock is faulty.
"Door is open (During the
cycle)"
This message is displayed when the
door is open: During the cycle.
"Canceled By User"
This message is displayed after the
STOP button is pressed and cycle
aborted.
Page 72
Summary of Contents for AMS10-120-T
Page 2: ......
Page 4: ...https www tuttnauerusa com...
Page 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Page 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Page 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Page 85: ...Page 78...