
Substance
Feed Water
Condensate
Chloride (Cl)
≤ 2 mg/l
≤ 0.1 mg/l
Phosphate
≤ 0.5 mg/l
≤ 0.1 mg/l
Conductivity (at
20°C)
≤15 µs/cm
≤ 3 µs/cm
pH value
5 to 7.5
5 to 7
Hardness
≤ 0.02 mmol/l
≤ 0.02 mmol/l
Appearance
Colorless, clean, without sediments
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Caution:
The use of water for autoclaves that do not comply with the table
above may have severe impact on the working life of the sterilizer
and can invalidate the manufacturer’s guarantee.
Use only deionized water, having a maximum conductivity of 15
µs/cm. Conductivity greater than 15 µs/cm may cause failures.
1.12.1
Tap water supply
The range of hardness value 0.7-2.0 mmol/l (70- 200 mg/l CaCO3)
The use of soft water is strictly forbidden!
Please consult a water specialist!
Note:
We recommend testing the water quality once a month. The use of water
for autoclaves that does not comply with the table above may have severe
impact on the working life of the sterilizer and can invalidate the
manufacturer’s guarantee.
Page 14
Summary of Contents for AMS10-120-T
Page 2: ......
Page 4: ...https www tuttnauerusa com...
Page 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Page 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Page 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Page 85: ...Page 78...