Caution!
In order to avoid any injury by electrical hazard, it is recommended
that a ground fault protection device (GFCI) be installed in the
electrical panel feeding the autoclave (local codes may make this
mandatory).
The
electrical network
must comply
with
local rules
and
regulations.
Verify that there is an easy access to the main power switch and to
the current leakage safety relay (GFCI). The voltage supplied to the
device must comply with the label ± 10%.
1.11
Device Environmental Information
l
The peak sound level generated by the autoclave is 67dBa with
background noise of 48dBa during sterilization stage, and 65dBa during
drying stage.
l
The total heat per hour transmitted by the autoclave is <200Wh.
1.12
Requirements with concern to Water Quality
The distilled or mineral-free water supply to the autoclave shall be according
to the table below:
Suggested Maximum Limits of Contaminants in Water for Steam
Sterilization per EN13060
Table 5 - Water Quality
Substance
Feed Water
Condensate
Evaporate residue
≤ 10 mg/l
≤ 1.0 mg/l
SiO
2
≤ 1 mg/l
≤ 0.1 mg/l
Iron
≤ 0.2mg/l
≤ 0.1mg/l
Cadmium
≤ 0.005 mg/l
≤ 0.005 mg/l
Lead
≤ 0.05 mg/l
≤ 0.05 mg/l
Rest of heavy metals
except iron,
cadmium, lead
≤ 0.1 mg/l
≤ 0.1 mg/l
Page 13
Summary of Contents for AMS10-120-T
Page 2: ......
Page 4: ...https www tuttnauerusa com...
Page 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Page 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Page 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Page 85: ...Page 78...