8.4
Virus Protect program
The
Virus Protect
is a program that eliminates viruses and is performed
prior to a sterilization cycle.
Note: The Virus Protect is not for use for Textile and Porous programs.
1. Select from a sterilization cycle, the
Virus Protect
option.
2. Raise the temperature to 230°F (110°C) and wait for a period of 10
minutes, for the Virus Protect process to complete and if, for example,
the Unwrapped Instr. sterilization program is selected, the screen below
is displayed if the virus protect is successfully completed.
Page 49