Message
Description
Please fill water tank to full
for start
For prion cycle in manual water filling
configuration, when the water tank is not
full the “Start” button will not appear,
instead this error message will appear in
the Active alarms screen.
Prion cycle, in manual water filling and
the water tank is not full
Unrecognized printer
If printer is not recognized an error will
appear in the UI.
Low Quality Water
1. Read the water level from analog
sensor.
2. Translate the digital value to micro
siemens according to the attached
table.
3. Compare
the
value
system
parameter
(water
quality
reference, default 10).
4. if value is less than 10 show error
- Low Quality Water
Set Atmosphere Pressure
is Active
Present when Reset Atmospheric
pressure is requested from Technician
screen
“No Water”
Chamber clean cycle, If Float switch (DI
2) doesn’t change from 1 to 0 during
water pump time On.
“Analog Input Error -
Jacket Temperature”
Utility issue #1 – Please
switch OFF and ON
machine’s power switch
If windows card disconnects
Utility issue #2 – Please
switch OFF and ON
If WIN application crash/not start up