Message
Description
This message and symbol are displayed
if an error occurs before sterilization
cycle is completed.
"Air Error"
This message is displayed at the end of
the cycle if the autoclave does not reach
the atmospheric pressure after 10
minutes.
If the water level is low in the clean water
reservoir and not sufficient for at least
one cycle, this symbol will be displayed.
Mineral free water reservoir empty will
appear in the error messages.
"Routine cycle service is
recommended Please call
your service provider."
Number of cycles, since last periodical
maintenance, exceeded the "cycle
service counter" parameter Or time
elapsed exceeded the "time service
counter" parameter.
"Power Down"
This message is displayed if power down
has occurred during the cycle. (this
message will print out in the printer after
the autoclave will turn on).
Vacuum time error”
Vacuum pressure fails to reach the
required value within the required time
(in Vacuum test and Pre-vacuum stage)
Water fill Error
If DO10 is On for timer > Parameter
Water fill time out and DI2 (Max. float
switch) is not On
Water tank filling, Please
wait
If high water demand cycle is chosen
(e.g. Prion) and Parameter>0 (Auto fill) -
Then before enabling Start message
presented on GUI
Page 73
Summary of Contents for AMS10-120-T
Page 2: ......
Page 4: ...https www tuttnauerusa com...
Page 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Page 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Page 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Page 85: ...Page 78...