MAN205-0195042EN Rev. A
Aug. 2011
Page 84 of 137 pages
9.2
Safety Tests after Repair
Attention!
After every repair or dismantling the enclosure, the autoclave should pass two
electrical safety test by the qualified Service Engineer. The following shall be
performed:
1.
Enclosure Leakage Current Test.
Every autoclave should pass this test as follows:
1.
Connect the electrical cord to the autoclave.
2.
Turn ON the main switch on the right side of the autoclave.
3.
Short-circuit the L and N pins on the cord's plug.
4.
Connect the Short-circuit pins to the L pole on the Megaohm Tester.
5.
Connect the earth pins to the earth pole on the Megaohm Tester.
6.
Impose an electrical potential of 500
–1000V on the tested autoclave.
The insulation resistance should be at le
ast 2 MΩ.
The test is successful if there was no leakage.
2.
Protective Earth Impedance Test
1.
Connect the grounding pin of the power cord plug to one pole of an
Ohmmeter.
2.
Connect any other metallic part (preferable
– the metallic part of the
locking screw) to the second pole of the Ohmmeter.
3.
The resistance should not exceed 0.3 Ω.
After performing these tests, the Service Engineer should complete and sign the
Work Order and the Log Book.
Warning!
It is strictly forbidden for any person to enter the service area behind the service
access panels, except the trained technician.
If, for any reason it is necessary to open the service access panel, the person
must shut the system by pressing the emergency switch, withdrawing the key and
keeping the key on him, to prevent accidents and injuries.
The technician must follow the instructions in the SERVICE INSTRUCTIONS,
“Working with power ON in Service Area” Section.
Summary of Contents for 6671130
Page 2: ...Web site http www tuttnauer com...
Page 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Page 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Page 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Page 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Page 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...