MAN205-0195042EN Rev. A
Aug. 2011
Page 97 of 137 pages
Replacing the Water Level Electrodes
1.
Turn OFF the electric power and discharge the steam pressure from the
autoclave and the steam generator.
2.
Disconnect the wire from the electrode connections.
Note:
Mark the wires before removing them. It is essential to return the correct wire to
the correct electrode!
3.
Remove the electrode.
4.
Insert a new electrode into their place.
5.
Set the height as per the drawing on the below, and tighten the fixing
nut to avoid leakage.
Note:
The setting of the electrode height is essential the proper working of the
generator!
6.
Reconnect the wire to the electrodes Return the correct wire to the
correct electrode.
7.
Test the unit.
Dimensions in mm
Front of Generator
Summary of Contents for 6671130
Page 2: ...Web site http www tuttnauer com...
Page 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Page 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Page 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Page 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Page 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...