9.8
Releasing the Vacuum Pump Rotor
The vacuum pump could get locked after a prolonged idle period.
Further described, is the procedure to follow in order to unlock and restore the normal
operation of the pump.
1.
When the pump is operated for the first time, check if it works normally.
If not, stop the electric current immediately.
2.
Do not open the pump and by no means, try to free the rotor shaft, by
forcibly rotating the impeller, this can provoke an irreversible damage to
the pump.
3.
Further described is the way to proceed, for restoring the normal
operation of the pump.
a.
By means of a screwdriver, rotate in clockwise direction the shaft of
the pump on which the cooling fan of the motor is mounted.
The shaft will be unlocked and the pump is ready for operation.
b.
If it did not help, unscrew the pipes connected to the pump and pour
in a quantity of 200ml. of edible vinegar available at any grocery
shop.
Leave the pump in this situation for ½ - 1 hour.
c.
By means of a screwdriver, rotate the shaft of the pump on which the
cooling fan of the motor is mounted in a clockwise direction. The
shaft will be unlocked and the pump is ready for operation.
Summary of Contents for 6671130
Page 2: ...Web site http www tuttnauer com...
Page 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Page 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Page 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Page 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Page 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...