MAN205-0195042EN Rev. A
Aug. 2011
Page 17 of 137 pages
3.1.1
Electrical Safety System
There is an electrical safety system to protect the heating elements if a fault condition is
detected by any one of the safety devices, namely:
Water level too low
Pressure too high
Temperature too high, (if applicable)
The fault light on the generator box will light up, and the main heater circuit breakers will
be switched OFF.
Description:
Page:
Design:
Rev No.
Drawing:
Prepared by:
Job No.
Approved by:
Scale:
Quant./Job:
Drawing No.
Cat. No.
Q. C.
Quant./Unit:
01
ISO
A3
Preparation By:
Description
Date
Rev. No
ECO
Generator electrical layout for
2 sets of heater with single phase
From:
1
4
ES1G08-0001-SP
Marchel N.
22.8.10
Yuval S.
22.8.10
See page 4
400 mm
200 mm
4
0
0
m
m
Main Switch
4
2
4
1
4
0
3
9
3
8
3
7
3
6
3
5
4
6
4
5
4
4
4
3
TB4
7
1
PS
24VDC
8
NO
CR1
5
1
L1
3
L3
13
NO
5
L5
2
T2
4
T4
6
T6
14
NO
A1
A2
Heater 2
KM2
3
1
L1
3
L3
13
NO
5
L5
2
T2
4
T4
6
T6
14
NO
A1
A2
Heater 1
KM1
3
PLC
CB
L1
T1
CB3
L1
T1
CB4
4
4
L3
T3
L2
T2
T1
L1
CB1
3
L3
T3
L2
T2
T1
L1
CB2
3
TC Test
3ph Line Feed
PLC
CB
D1
6
D2
TC2
PLC
CB
D1
6
D2
TC1
9
R/L1
S/L2
T/L3
Power
Heaters
Water
Fault
GICAR
2DRLH
S
R
1
S
M
1
S
m
1
S
R
2
S
M
2
S
m
2
RL1
RL2
S
E
S
E
L N
R
L
1
R
L
2
10
To Water
Pump
PE
TB3
L1
TB2
L2 L3
3 ph
To control panel
L
L
PE
TB5
N
1 ph
To control panel
N
PE
TB6
L
N
1 ph
Feeding
PE
N
TB1
11
If this happens, do not try to correct the fault yourself, but call for assistance.
3.1.2
The Water Pump for the Steam Generator
The water pump supplies the distilled or mineral-free water to the steam generator.
The supply line of the pump is connected to the upper water reservoir filled with distilled
or mineral-free water and mounted on side of the autoclave.
A float valve keeps a constant level of water and prevents an overflow of water out of the
reservoir.
A lower float switch stops the operation of pump, when the water level drops below the
lower level limit.
Fault Light
On Generator
electric box
Circuit Breaker
(OFF)
Trip Coil
Summary of Contents for 6671130
Page 2: ...Web site http www tuttnauer com...
Page 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Page 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Page 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Page 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Page 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...